Related Experiment Video
Updated: Feb 10, 2026

Single Port Donor Nephrectomy
Published on: March 12, 2011
Efficacy of vaporized hydrogen peroxide for repeated sterilization of a single-use single-incision laparoscopic
Valery F Scharf1, Brian Dent1, Megan E Jacob1
1College of Veterinary Medicine, North Carolina State University, Raleigh, North Carolina.
Objective:
To determine the ability of vaporized hydrogen peroxide (VHP) to sterilize a single-use single-incision laparoscopic surgery port and its associated components after repeated simulated uses.
Study Design:
Prospective in vitro experimental study.
Sample Population:
Six single-use single-incision laparoscopic surgery ports with associated cannulas and insufflation tubing.
Methods:
Ports, cannulas, and tubing were subjected to 10 cycles of simulated use, bacterial inoculation with Staphylococcus pseudintermedius and Escherichia coli, decontamination and sterilization, and testing via culture based on their treatment group designation. Bacteriological scores were compared among the negative control, positive control, and 4 treated ports and components.
Results:
There was no difference in bacteriological scores between treated ports, cannulas, and insufflation tubing and the negative control port and components. Bacteriological scores of ports and components undergoing 6-10 cycles were not significantly different from scores of ports and components undergoing 5 or fewer sterilization cycles. No difference was found in detection of bacteria from treated ports by biopsy of the foam versus sampling via wash.
Conclusion:
This study suggests that a single-use single-incision laparoscopic port and its associated components can be effectively sterilized after multiple simulated uses by using VHP.
Clinical Significance:
Reuse of a single-use single-incision laparoscopic port is a safe and effective method of cost reduction in veterinary patients.
More Related Videos
03:32Single Incision Plus One Port Laparoscopic Proximal Gastrectomy with Double Channel Anastomosis for Gastric Cancer Treatment
Published on: December 27, 2024
03:07Single-Port Robotic-assisted Transaxillary Breast-conserving Surgery: A Prospective, Single-arm, Non-randomized Phase IIa Clinical Trial
Published on: August 19, 2025
Related Concept Videos
Vapor Pressure
Vaporization
Phase Transitions: Vaporization and Condensation
Vapor Pressure Lowering
Hydrogen Bonds
Hydrogen Bonds Control the World!
Because hydrogen has very weak electronegativity when it binds with a strongly electronegative atom, such as oxygen or nitrogen, electrons in the bond are unequally shared....
Load along a Single Axis
Consider a beam of length L subjected to a varying load, which is a combination of parabolic and trapezoidal load distribution along the x-axis. In this case, it is essential to determine the resultant loads, their locations, and...