Qualitative study of mechanical parameters of conventional diagnostic X-ray machines in Mizoram
Jonathan Lalrinmawia1, Kham Suan Pau2, Ramesh Chandra Tiwari3
1Department of Physics, Mizoram University, Aizawl, Mizoram, 796004, India.
Abstract:
The present study examined the mechanical attributes of 135 conventional diagnostic X-ray machines in Mizoram, India. The purpose of studying the X-ray mechanical parameters, such as congruency, perpendicularity of the central beam, and half-value layer, was to improve the quality of the diagnostic image and reduce the patient dose. A battery-operated portable dosimeter was used to measure output radiation of the X-ray machine. The half-value layer was measured at a constant accelerating potential of 70 kVp and tube load. To measure the congruency and beam alignment perpendicularity, a congruence and alignment tool was used. The survey data were collected between June 2015 and June 2016. The authors followed international standard test procedures, and the results were compared to national and international standards. SPSS Statistics for Windows, Version 17 was used to calculate the mean, range, and standard deviation. The half-value layer ranged from 0.45 to 3.00 mm; the mean half-value layer was 1.60 ± 0.51 SD mm. In comparison with national and international standards, only 27.83% (national) and 15.64% (international) of the machines' filtration were found to be within acceptable limits. The congruence misalignment of the x-axis varied between 0.50% and 15.30% of the source-to-image distance; for the y-axis, it ranged from 0.50 to 10.90%. When the congruence between the radiation beam and optical field was tested, 80.85% of diagnostic X-ray machines did not meet the prescribed acceptance parameters. When the perpendicularity between the central beam and the image receptor was tested, 69.81% did not meet safety standards.
Related Concept Videos
Enolate Mechanism Conventions
Mechanical Efficiency of Real Machines
However, in reality, no machine can be truly ideal, and all of them experience some...
X-ray Crystallography
Diffraction
Diffraction is the change in the direction of travel experienced by an electromagnetic wave when it encounters a physical barrier whose dimensions are comparable to those of the wavelength of the light. X-rays are electromagnetic radiation with wavelengths about as long as the distance between neighboring...
Qualitative Analysis
For instance, group IV...
Qualitative Analysis
There are two main approaches to qualitative analysis:...
Appendicitis-II: Diagnostic Studies and Management
Diagnosing Appendicitis
It requires a multifaceted approach, starting with a detailed physical examination to pinpoint the location and nature of the pain and identify any associated symptoms. Laboratory tests play a crucial role. A complete Blood Count (CBC) typically reveals leukocytosis (an increased number of...