Peripheral Intravenous Volume Analysis (PIVA) for Quantitating Volume Overload in Patients Hospitalized With Acute
Merrick Miles1, Bret D Alvis1, Kyle Hocking2
1Vanderbilt University Medical Center, Department of Anesthesiology, Division of Critical Care, Nashville, Tennessee.
Background:
To determine the feasibility of peripheral intravenous volume analysis (PIVA) of venous waveforms for assessing volume overload in patients admitted to the hospital with acute decompensated heart failure (ADHF).
Methods:
Venous waveforms were captured from a peripheral intravenous catheter in subjects admitted for ADHF and healthy age-matched controls. Admission PIVA signal, brain natriuretic peptide, and chest radiographic measurements were related to the net volume removed during diuresis.
Results:
ADHF patients had a significantly greater PIVA signal on admission compared with the control group (P = .0013, n = 18). At discharge, ADHF patients had a PIVA signal similar to the control group. PIVA signal, not brain natriuretic peptide or chest radiographic measures, accurately predicted the amount of volume removed during diuresis (R2 = 0.781, n = 14). PIVA signal at time of discharge greater than 0.20, demonstrated 83.3% 120-day readmission rate.
Conclusions:
This study demonstrates the feasibility of PIVA for assessment of volume overload in patients admitted to the hospital with ADHF.
More Related Videos
05:16Cutoff Value of Phase Angle by Bioelectrical Impedance Analysis at Admission as a Prognostic Factor in Patients with Acute Heart Failure
Published on: June 10, 2025
06:04Author Spotlight: Establishment and Confirmation of a Postnatal Right Ventricular Volume Overload Mouse Model
Published on: June 9, 2023
Related Concept Videos
Constant Volume Calorimetry
Volume of Distribution
Problem Solving: Volume
Respiratory Volumes
Tidal Volume (TV) Tidal volume (TV) is the air inhaled or exhaled in a...
Work Done During Volume Change
Consider a gas confined to a cylinder fitted with a movable piston at one end. If the gas expands from volume V1 to volume V2, it exerts a force on the piston, such that the piston moves by a distance dr.
The work done by the gas on the piston can be expressed as
Control Volume and System Representations
The control volume approach considers a stationary region in space through which fluid flows. This region is bounded by a control surface. For instance, in the case of water...