Effect of Repeated Sterilization on the Tensile Strength of Rubber Dam Clamps
Mudrika Chhabra1, Ann L Greenwell2
1Private practice, Tallahassee, Fla., USA;,
Purpose:
The fracture of rubber dam clamps after repeated clinical uses can result in ingestion/aspiration of the fragments. This pilot study was aimed at examining whether the tensile strength of a rubber dam clamp changes significantly following the stresses of repeated clinical use and steam sterilizations.
Methods:
One-hundred new stainless-steel rubber dam clamps were subjected to repeated steam sterilization and placement on a wooden dowel periodically, simulating clinical use. Two clamps were randomly selected for removal after every 50 sterilization cycles. Three-hundred such cycles were completed. Using a standard bench top universal tensile testing system (MTI-2K) a tensile force was applied to break the new clamps and the clamps that were stretched and autoclaved. The maximum force was recorded, and data analyzed using simple mean. The mean tensile load values were plotted on a graph against the number of sterilization cycles to analyze trends in the data.
Results:
The rubber dam clamps demonstrated a higher tensile strength and fracture resistance after three-hundred cycles of use and sterilization.
Conclusion:
Repeated use and steam sterilization up to three-hundred cycles did not adversely affect the tensile strength of the rubber dam clamps. The clamps, in fact, demonstrated a higher tensile strength.
Related Concept Videos
Tensile Strength Considerations of Concrete
The dimensions and shape of a concrete specimen...
Relation Between Tensile Strength and Compressive Strength of Concrete
Acid Strength and Molecular Structure
In the absence of any leveling effect, the acid strength of binary compounds of hydrogen with nonmetals (A) increases as the H-A bond strength decreases down a group in the periodic table. For group 17, the order of increasing acidity is HF < HCl < HBr < HI. Likewise, for group 16, the order of increasing acid strength is H2O < H2S < H2Se < H2Te. Across a row in the periodic table, the acid strength of binary hydrogen compounds increases with increasing...
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Strength of Cement
For compressive strength tests, ASTM C 109-05 standards prescribe a cement-sand mix ratio of 1:2.75 and a water/cement ratio of 0.485 for making 2-inch cubes. These cubes are mixed, cast, and cured in saturated lime water at 23°C until testing. Flexural strength testing, outlined in...