Performance of a novel, manually operated intraosseous device in adult human cadavers
Sandeep Singh1, Balram Bhargava1, Praveen Aggarwal2
1Department of Cardiology, All India Institute of Medical Sciences, New Delhi, India.
Aim:
Intraosseous (IO) access in adults is preferred using semi-automatic devices as it is difficult to penetrate the thick cortical layer of long bones using manual needles. The authors have developed an IO device which generates both rotational and axial thrust using a manual driver. This drilling mechanism addresses certain pain-points of current IO devices. The objective of this study was to evaluate the performance of this device in human cadavers.
Methods:
The authors tested the ability of this device for IO access at proximal and distal tibia in 10 adult cadavers. Needle position was confirmed by fluoroscopy after contrast injection. Needle penetration time-defined as the time required for manual drilling of bone-and the total procedure time were calculated from video analysis. A successful IO procedure was defined as an appropriate needle position without any contrast extravasation, device, or procedure-related complication. After each procedure, the authors recorded damage to the device or fracture of the bone.
Results:
A single physician performed 40 IO procedures. The IO access was successful in 35 (87.5 percent) and was accomplished in first attempt in 33 (82.5 percent) insertions. Reasons for failure were undershooting of needle (2/40, 5 percent), trocar damage (1/40, 2.5 percent), and detachment of plastic hub of the needle during removal in (2/40, 5 percent) procedures. There were no bone fractures. In all but one instance, needle penetration time was <3 seconds. The mean total procedure time was 40 ± 13 seconds.
Conclusion:
In this pilot study, the authors have demonstrated the efficacy of a novel, manually introduced IO device in adult cadavers.
Related Concept Videos
Real Number Operations
Operational Amplifiers
Vector Operations
A vector multiplied by a scalar value is called scalar multiplication. The result obtained is a new vector with a different magnitude. If the scalar is positive, the direction of the vector remains the same, but if it is negative, the direction of the vector is reversed. For example, the product of the mass and velocity yields the momentum.
Operant Conditioning
Reinforcement in operant conditioning can be positive or negative, both of which serve to increase the likelihood of a behavior. Positive...
Second Derivatives and Laplace Operator
Consider a scalar function. The curl of its...
Modes of Operations of BJT
Active Mode: The most common mode for amplification, the active mode features a forward-biased emitter-base junction and a reverse-biased base-collector junction. This setup enables electrons to be injected from the emitter to the base while blocking the majority carriers at the collector. The...