Differences in corneo-scleral topographic profile between healthy and keratoconus corneas
David P Piñero1, Antonio Martínez-Abad2, Roberto Soto-Negro3
1Group of Optics and Visual Perception, Department of Optics, Pharmacology and Anatomy, University of Alicante, Spain; Department of Ophthalmology (OFTALMAR), Vithas Medimar International Hospital, Alicante, Spain.
Purpose:
To evaluate the differences in corneo-scleral topographic profile between healthy and keratoconus eyes, and their potential diagnostic ability for keratoconus detection.
Methods:
Prospective comparative study including 21 keratoconic eyes (11 patients) and 88 healthy eyes (88 patients). In all cases, a complete eye exam was performed including an evaluation of the corneo-scleral profile. The diagnostic ability of corneo-scleral topographic parameters to detect keratoconus was evaluated using the receiver operating characteristic (ROC) curve.
Results:
A significant lower inferior tangent angle at limbus (ITA) was found in the keratoconic group compared to the control group (p = 0.024). Regarding sagittal heights, significant differences between groups were found in temporal sagittal height (TSH) for 11 mm (p = 0.040), 12 mm (p = 0.041) and 13 mm corneal chords (p = 0.040), difference between temporal and nasal sagittal heights (T-NSH) for 12 mm (p = 0.025) and 13 mm (p = 0.034), and maximum sagittal height (MaxSH) for 12 mm (p = 0.043), with higher values in keratoconus. In bilateral cases, these differences were not found when comparing with the least severe keratoconus eye. Statistical significance for the ROC curve was only found for ITA (p = 0.025), 12-mm (p = 0.048) and 13-mm TSH (p = 0.042), and 13-mm T-NSH (p = 0.037), with cutoff values associated to limited values of sensitivity and specificity.
Conclusions:
The corneo-scleral profile in keratoconus presents higher levels of asymmetry compared to healthy eyes, especially in eyes with moderate and advanced stages of the disease. The diagnostic accuracy of corneo-scleral topographic data alone for keratoconus detection is limited and must be used in conjunction with other clinical parameters.
More Related Videos
14:58A Simplified Technique for In situ Excision of Cornea and Evisceration of Retinal Tissue from Human Ocular Globe
Published on: June 12, 2012
05:56Scleral Cross-linking Using Riboflavin and Ultraviolet-A Radiation for Prevention of Axial Myopia in a Rabbit Model
Published on: April 3, 2016
Related Concept Videos
Topographic Surveying and Contours
Plotting of Topographic Maps
Electric Potential and Potential Difference
When a test charge moves from the initial to the final position, the electric potential difference between those positions is defined as the ratio of the change in the potential energy to the charge on the...
Difference from Background: Limit of Detection
The LOD indicates the presence or absence...
Identifying Statistically Significant Differences: The F-Test
Sum and Difference OpAmps
A summing amplifier, or an adder, utilizes an op-amp to merge multiple input signals into a single output signal. When audio signals are introduced into its input channels, the input resistors initiate currents that traverse feedback resistors, resulting in an output voltage. Applying Kirchhoff's current...