In-vivo stiffness assessment of distal femur fracture locked plating constructs
Christopher Parks1, Christopher M McAndrew2, Amanda Spraggs-Hughes2
1EmergeOrtho, Wilmington, NC, USA.
Background:
The purpose of this study was to design and validate a novel stiffness-measuring device using locked plating of distal femur fractures as a model.
Methods:
All patients underwent a laterally-based approach, with a bridging locked construct after indirect reduction. A custom and calibrated intraoperative stiffness device was applied and the stiffness of the construct was blindly recorded. Fourteen of twenty-seven patients enrolled with distal femur fractures (AO/OTA 33A and 33C) completed the study. Correlations between stiffness and callus formation, working length, working length/plate length ratio, number of distal locking screws, and fracture pattern were explored.
Findings:
Callus and modified radiographic union scale in tibias scores as a linear function of stiffness did not correlate (R2 = 0.06 and 0.07, respectively). Construct working length and working length to plate length ratio did not correlate to stiffness (R2 = 0.18 and 0.16 respectively). A combined delayed and nonunion rate was 14%. Lower extremity measure scores were not statistically different when comparing delayed and nonunion with healed fractures.
Interpretation:
The lack of correlation may have been due to the mechanical properties of the plate itself and its large contribution to the overall stiffness of the construct. To our knowledge, clinically relevant stiffness has not been described and this study may provide some estimates. This methodology and these preliminary findings may lay the groundwork for further investigations into this prevalent clinical problem. Other parameters not investigated may play a key role such as body mass index and bone mineral density.
Level Of Evidence:
Diagnostic/Prognostic Level II.
Related Concept Videos
Self-Locking Screw
A square-threaded screw jack carrying a load is considered self-locking if the screw retains its position even after the moment applied to it is removed.
Bones of the Lower Limb: Femur and Patella
Fractures: Bone Repair
Minor fractures with no bone displacement are treated by immobilizing the fractured bone using a cast or splint. However, in the case of fractures with displaced bones, the broken bones are repositioned before immobilization to ensure successful healing without deformation and loss of function. The realignment of fractured bone ends is performed through a process called reduction. If the...
Bode Plots Construction
Construction of Root Locus
For positive gain values, the root locus exists on the real axis to the left of an odd number of finite open-loop poles or zeros. The root locus starts at the open-loop poles and traces the paths of the closed-loop poles as the gain...
Construction of Frequency Distribution
First, make a table with two columns—one with the title of the data that needs to be organized, and the other column for frequency. [Draw a third column for tally marks if needed]. Then, take a look at the items given in the data set and decide if an ungrouped frequency distribution table or a grouped frequency distribution table would be more suitable. If there are large sets of different values, then it is...