Retrieval of a Micra transcatheter pacing system in a heart with a preexisting lead
Junji Morita1, Masato Fukunaga1, Kenichi Hiroshima1
1Department of Cardiology, Kokura Memorial Hospital, Japan.
Abstract:
We report the case of a 74-year-old man with a previously implanted pacemaker lead. He had undergone Medtronic™ Micra Transcatheter Pacing System (TPS, Medtronic plc, MN, USA) implantation because of lead fracture. We implanted a new TPS and retrieved the dislodged one. We used a multiple-loop snare (EN snare®) and an 8.5F steerable sheath (Agilis NXT; St. Jude Medical, St Paul, MN, USA). The TPS was obstructed by the chordae tendineae of the tricuspid valve and the pacemaker lead. We pushed the TPS to the apex site; this enabled us to move the TPS away from the chordae tendineae and pacemaker lead. The TPS body was caught in the inferior vena cava and was successfully retrieved. To our knowledge, this is the first case reporting TPS retrieval in a heart with preexisting lead.
More Related Videos
Related Concept Videos
Retrieval
Recall involves accessing information without cues, such as during an essay test, where individuals must retrieve facts and concepts from memory unaided. Another example is remembering the name of a colleague...
ER Retrieval Pathway
The ER uses many checkpoints to prevent the entry of incorrectly folded or a resident protein as cargo onto a transport vesicle. These mechanisms...
Anatomy of the Heart
Anatomy of the Heart
The heart has three layers: the innermost endocardium, the muscular myocardium, and the outer epicardium, all working together for optimal cardiac function.
Chambers of the Heart
The heart is made up of four...
Phase-lead and Phase-lag Controllers
Time and frequency -Domain Interpretation of Phase-lead Control
The design of phase-lead control involves the strategic placement of poles and zeros to balance steady-state error and system...