Type testing of a locally made LiF:Mg,Ti + PTFE TLD for its use as a personal dosimeter
Juan Azorín1, Eugenio Torijano1, Claudia Azorín2
1Universidad Autónoma Metropolitana, Unidad Iztapalapa, Mexico City, Mexico.
Abstract:
Our research group has developed a thermoluminescence dosimeter (TLD) based on pellets of LiF:Mg,Ti mixed with polytetrafluoroethylene (LiF:Mg,Ti +PTFE). This TLD can be used as a personal dosimeter. Extensive type testing, carried out with reference to the International Electrotechnical Commission (IEC) Standard, were performed for the purpose of accepting the LiF:Mg,Ti+PTFE as a personal TL dosimeter. Tests performed include repeatability, batch homogeneity, linearity, detection threshold, and light sensitivity. Results showed that locally made LiF:Mg,Ti+PTFE TLDs met all the standard requirements.
Related Concept Videos
Self-Report Tests of Personality
Types of Hypothesis Testing
When the null and alternative hypotheses are stated, it is observed that the null hypothesis is a neutral statement against which the alternative hypothesis is tested. The alternative hypothesis is a claim that instead has a certain direction. If the null hypothesis claims that p = 0.5, the alternative hypothesis would be an opposing statement to this and can be put either p > 0.5, p < 0.5, or p...