Jove
Visualize
Contact Us
JoVE
x logofacebook logolinkedin logoyoutube logo
ABOUT JoVE
OverviewLeadershipBlogJoVE Help Center
AUTHORS
Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
LIBRARIANS
TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
RESEARCH
JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
EDUCATION
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
Terms & Conditions of Use
Privacy Policy
Policies

Related Concept Videos

Design Consideration01:22

Design Consideration

583
Designing a structure involves a series of considerations, primarily the material's ultimate strength, calculated through tests that measure changes under increased force until the material reaches its breaking point or limit. The ultimate load, where the material breaks, is divided by its original cross-sectional area, resulting in the ultimate normal stress or strength. The ultimate shearing stress is another significant factor taken into account.
The factor of safety is another key...
583
Transmission Line Design Considerations01:23

Transmission Line Design Considerations

633
Aluminum has become the material of choice for overhead transmission lines, surpassing copper due to its abundance and cost-effectiveness. The most prevalent type is the aluminum conductor, steel-reinforced (ACSR), which combines aluminum strands around a steel core. Other variants include all-aluminum conductors (AAC), all-aluminum alloy conductors (AAAC), aluminum conductor alloy-reinforced (ACAR), and aluminum-clad steel conductors. Advanced designs, such as aluminum conductors with steel...
633
Physical and Chemical Properties of Matter02:57

Physical and Chemical Properties of Matter

166.9K
The characteristics that enable us to distinguish one substance from another are called properties.
166.9K
Group Design02:01

Group Design

10.7K
The most basic experimental design involves two groups: the experimental group and the control group. The two groups are designed to be the same except for one difference— experimental manipulation. The experimental group gets the experimental manipulation—that is, the treatment or variable being tested—and the control group does not. Since experimental manipulation is the only difference between the experimental and control groups, we can be sure that any differences between...
10.7K
Tensile Strength Considerations of Concrete01:16

Tensile Strength Considerations of Concrete

558
Considering the tensile strength of concrete involves recognizing that the theoretical strength of cement paste can be up to a thousand times higher than what is observed in practical applications. This significant discrepancy is largely attributed to the presence of microscopic cracks within the concrete. These cracks tend to amplify stress at their tips when a load is applied, a phenomenon explained by Griffith's theory of brittle fracture.
The dimensions and shape of a concrete specimen...
558
Physical Pendulum01:06

Physical Pendulum

2.8K
When a rigid body is hanging freely from a fixed pivot point and is displaced, it oscillates similar to a simple pendulum and is known as a physical pendulum. The period and angular frequency of a physical pendulum are obtained by using the small-angle approximation and drawing parallels with a spring-mass system. The small-angle approximation (sinθ=θ) is valid up to about 14°.
When dealing with complicated systems, the mass moment of inertia is an important parameter, as it...
2.8K

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

From Cellular Radiosensitivity to Precision Radiotherapy: Integrating Functional Assays, Genomics, and Clinical Modeling.

Cancers·2026
Same author

PRISM: An open-source framework for regularized material decomposition on a novel kV dual-layer imager.

Medical physics·2026
Same author

Radiation Oncology-Biology Integration Network: Bridging the Gap between Biological Research and Clinical Practice.

Clinical cancer research : an official journal of the American Association for Cancer Research·2026
Same author

Interobserver image registration variability impacts on stereotactic arrhythmia radioablation (STAR) target margins.

Biomedical physics & engineering express·2026
Same author

Patient outcome prognosis for external beam radiation therapy using CBCT-based radiomics: a systematic review.

Biomedical physics & engineering express·2025
Same author

CT online adaptive radiotherapy is associated with dosimetric and acute toxicity improvements in prostate cancer treatment.

Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology·2025
Same journal

Effective contrast-enhanced preprocessing for intracranial artery segmentation in digital subtraction angiography.

Physics in medicine and biology·2026
Same journal

Improving Plan Quality in Adaptive Proton Therapy Using an Interactive Dose Modification Tool.

Physics in medicine and biology·2026
Same journal

Technical Note: Real-Time MLC Control and Latency Measurement Optimization with External Verification.

Physics in medicine and biology·2026
Same journal

Fetus-Specific Hematopoietic Stem Cell Dosimetry Framework for Leukemia-Relevant Target Cells During Prenatal Development.

Physics in medicine and biology·2026
Same journal

Deep learning-based dose prediction to enhance planning efficiency in cervical brachytherapy with hybrid applicators.

Physics in medicine and biology·2026
Same journal

Corrigendum: Referenceless MR thermometry-a comparison of five methods (2017<i>Phys. Med. Biol</i>.<b>62</b>1-16).

Physics in medicine and biology·2026
See all related articles

Related Experiment Video

Updated: Feb 9, 2026

Synthetic, Multi-Layer, Self-Oscillating Vocal Fold Model Fabrication
10:16

Synthetic, Multi-Layer, Self-Oscillating Vocal Fold Model Fabrication

Published on: December 2, 2011

14.5K

Physics considerations in MV-CBCT multi-layer imager design.

Yue-Houng Hu1, Rony Fueglistaller2, Marios Myronakis1

  • 1Department of Radiation Oncology, Division of Medical Physics and Biophysics, Brigham and Women's Hospital, Dana-Farber Cancer Institute and Harvard Medical School, Boston, MA 02115, United States of America.

Physics in Medicine and Biology
|May 31, 2018
PubMed
Summary
This summary is machine-generated.

Advancing megavoltage cone-beam CT (CBCT) with electronic portal imaging devices (EPIDs) requires improved image quality and reduced dose. Multi-layer imagers (MLIs) show promise, with performance improving linearly with the number of layers, primarily limited by detector noise.

More Related Videos

Author Spotlight: Real-Time Imaging of Bonding in 3D-Printed Layers
04:36

Author Spotlight: Real-Time Imaging of Bonding in 3D-Printed Layers

Published on: September 1, 2023

3.9K
Multi-layer Cortical Ca2+ Imaging in Freely Moving Mice with Prism Probes and Miniaturized Fluorescence Microscopy
10:35

Multi-layer Cortical Ca2+ Imaging in Freely Moving Mice with Prism Probes and Miniaturized Fluorescence Microscopy

Published on: June 13, 2017

32.2K

Related Experiment Videos

Last Updated: Feb 9, 2026

Synthetic, Multi-Layer, Self-Oscillating Vocal Fold Model Fabrication
10:16

Synthetic, Multi-Layer, Self-Oscillating Vocal Fold Model Fabrication

Published on: December 2, 2011

14.5K
Author Spotlight: Real-Time Imaging of Bonding in 3D-Printed Layers
04:36

Author Spotlight: Real-Time Imaging of Bonding in 3D-Printed Layers

Published on: September 1, 2023

3.9K
Multi-layer Cortical Ca2+ Imaging in Freely Moving Mice with Prism Probes and Miniaturized Fluorescence Microscopy
10:35

Multi-layer Cortical Ca2+ Imaging in Freely Moving Mice with Prism Probes and Miniaturized Fluorescence Microscopy

Published on: June 13, 2017

32.2K

Area of Science:

  • Medical Physics
  • Radiological Imaging
  • Radiation Oncology

Background:

  • Megavoltage (MV) cone-beam computed tomography (CBCT) using electronic portal imaging devices (EPIDs) offers advanced capabilities for radiation therapy.
  • Widespread adoption is hindered by limitations in image quality and concerns regarding imaging radiation dose.
  • Innovations in detector design, including multi-layer imager (MLI) architectures and novel scintillation materials, are crucial for overcoming these challenges.

Purpose of the Study:

  • To investigate the impact of detector design elements, specifically multi-layer architecture and scintillation materials, on EPID-based CBCT image quality and object detectability.
  • To evaluate image quality metrics such as modulation transfer function (MTF) and noise power spectrum (NPS) from projection to 3D reconstructed images.
  • To assess object detectability using the ideal observer signal-to-noise ratio (d') as a figure-of-merit.

Main Methods:

  • Utilized a hybrid Monte Carlo and linear systems model to simulate detector performance.
  • Analyzed image quality metrics (MTF, NPS) across different stages of image reconstruction.
  • Applied a task-based figure-of-merit (d') to quantify the effect of detector design on object detectability.

Main Results:

  • Object detectability was generally limited by detector noise performance.
  • Implementing MLI imagers with a single scintillation material per layer demonstrated improved noise performance and d' that scaled linearly with the number of layers.
  • Thicker scintillators improved X-ray absorption and DQE(0), but low light yield could lead to electronic noise dominance and rapid image quality degradation at higher doses.
  • Optimizing heterogeneous MLI detectors prioritized minimizing total noise, with spatial resolution being a secondary consideration influenced by the intensity contribution and MTF of individual layers.

Conclusions:

  • Multi-layer imager design, particularly the number of layers and material selection, significantly impacts EPID-based CBCT image quality and object detectability.
  • Detector noise is a primary limiting factor for object detection in MV CBCT.
  • Balancing X-ray absorption, light yield, and spatial resolution across multiple layers is essential for maximizing the performance of advanced EPID detector designs.