Screening for obstructive sleep apnea among hospital outpatients
Michel Hug1, Katrin Uehli2, Stig Solbach1
1Faculty of Medicine, University of Basel, Basel, Switzerland.
Background:
Obstructive sleep apnea syndrome (OSAS) is common in adults. People with OSAS have a higher risk of experiencing traffic accidents and occupational injuries (OIs). We aimed to clarify the diagnostic performance of a three-channel screening device (ApneaLinkTM) compared with the gold standard of full-night attended polysomnography (PSG) among hospital outpatients not referred for sleep-related symptoms. Furthermore, we aimed to determine whether manual revision of the ApneaLinkTM autoscore enhanced diagnostic performance.
Methods:
We investigated 68 patients with OI and 44 without OI recruited from the University Hospital Basel emergency room, using a cross-sectional study design. Participating patients spent one night at home with ApneaLinkTM and within 2 weeks slept for one night at the sleep laboratory. We reanalyzed all ApneaLinkTM data after manual revision.
Results:
We identified significant correlations between the ApneaLinkTM apnea-hypopnea index (AHI) autoscore and the AHI derived by PSG (r = 0.525; p <0.001) and between the ApneaLinkTM oxygen desaturation index (ODI) autoscore and that derived by PSG (r = 0.722; p <0.001). The ApneaLinkTM autoscore showed a sensitivity and specificity of 82% when comparing AHI ≥5 with the cutoff for AHI and/or ODI ≥15 from PSG. In Bland Altman plots the mean difference between ApneaLinkTM AHI autoscore and PSG was 2.75 with SD ± 8.80 (β = 0.034), and between ApneaLinkTM AHI revised score and PSG -1.50 with SD ± 9.28 (β = 0.060).
Conclusions:
The ApneaLinkTM autoscore demonstrated good sensitivity and specificity compared with the gold standard (full-night attended PSG). However, Bland Altman plots revealed substantial fluctuations between PSG and ApneaLinkTM AHI autoscore respectively manually revised score. This spread for the AHI from a clinical perspective is large, and therefore the results have to be interpreted with caution. Furthermore, our findings suggest that there is no clinical benefit in manually revising the ApneaLinkTM autoscore.
Related Concept Videos
Sleep Apnea
The condition is more prevalent among...
Hospitals-II
Nurses that work in...
Hospitals-I
Insufficient Sleep and Sleep Deprivation
Sleep deprivation is a more severe form of sleep loss...
Stages of Sleep
Before sleep begins, in wakefulness, the brain exhibits primarily beta waves, which are high in frequency and low in amplitude, indicating alertness...
Understanding Sleep
The circadian rhythm, a nearly 24-hour cycle, is deeply influenced by environmental light cues. Light exposure directly affects the hypothalamus, which in turn regulates...