Comparison of Cliff-Lorimer-Based Methods of Scanning Transmission Electron Microscopy (STEM) Quantitative X-Ray
Andrea Parisini1, Stefano Frabboni2, Gian Carlo Gazzadi3
11CNR-IMM Sezione di Bologna,via P. Gobetti 101,40129 Bologna,Italy.
Abstract:
In this work, we compare the results of different Cliff-Lorimer (Cliff & Lorimer 1975) based methods in the case of a quantitative energy dispersive spectrometry investigation of light elements in ternary C-O-Si thin films. To determine the Cliff-Lorimer (C-L) k-factors, we fabricated, by focused ion beam, a standard consisting of a wedge lamella with a truncated tip, composed of two parallel SiO2 and 4H-SiC stripes. In 4H-SiC, it was not possible to obtain reliable k-factors from standard extrapolation methods owing to the strong CK-photon absorption. To overcome this problem, an extrapolation method exploiting the shape of the truncated tip of the lamella is proposed herein. The k-factors thus determined, were then used in an application of the C-L quantification procedure to a defect found at the SiO2/4H-SiC interface in the channel region of a metal-oxide field-effect-transistor device. As in this procedure, the sample thickness is required, a method to determine this quantity from the averaged and normalized scanning transmission electron microscopy intensity is also detailed. Monte Carlo simulations were used to investigate the discrepancy between experimental and theoretical k-factors and to bridge the gap between the k-factor and the Watanabe and Williams ζ-factor methods (Watanabe & Williams, 2006).
More Related Videos
07:24Quantitative Atomic-Site Analysis of Functional Dopants/Point Defects in Crystalline Materials by Electron-Channeling-Enhanced Microanalysis
Published on: May 10, 2021
14:55Analysis of Minerals Produced by hFOB 1.19 and Saos-2 Cells Using Transmission Electron Microscopy with Energy Dispersive X-ray Microanalysis
Published on: June 24, 2018
Related Concept Videos
Transmission Electron Microscopy
Scanning Electron Microscopy
Fundamental Principles
Accelerated...
X-ray Crystallography
Diffraction
Diffraction is the change in the direction of travel experienced by an electromagnetic wave when it encounters a physical barrier whose dimensions are comparable to those of the wavelength of the light. X-rays are electromagnetic radiation with wavelengths about as long as the distance between neighboring...
Electron Affinity
Overview of Electron Microscopy
Immunogold Electron Microscopy