Refractive outcomes of intraocular lens power calculation using different corneal power measurements with a new
Giacomo Savini1, Kazuno Negishi1, Kenneth J Hoffer1
1From the G.B. Bietti Foundation IRCCS (Savini, Schiano Lomoriello), Rome, Italy; the Department of Ophthalmology (Negishi), Keio University School of Medicine, Hospital, Tokyo, Japan; Stein Eye Institute (Hoffer), University of California, Los Angeles, and St. Mary's Eye Center (Hoffer), Santa Monica, California, USA.
Purpose:
To evaluate the results of intraocular lens (IOL) power calculation using different corneal power measurements provided by an optical biometer combined with a dual Scheimpflug analyzer and a Placido disk topographer (Galilei G6).
Setting:
G.B. Bietti Foundation, Rome, Italy, and Keio University Hospital, Tokyo, Japan.
Design:
Evaluation of diagnostic technology.
Methods:
Consecutive patients having cataract surgery were enrolled. The IOL power was calculated with the Barrett Universal II, Haigis, Hoffer Q, Holladay 1, and SRK/T formulas. Different options were used to calculate the corneal power: simulated keratometry (K) based on anterior corneal surface measurements only and total corneal power (TCP) based on ray tracing through both corneal surfaces. Three TCP measurements (TCP1, TCP2, and TCP-IOL) were evaluated.
Results:
The study analyzed 118 eyes. The mean values of simulated K (43.74 diopters [D] ± 1.40 [SD]), TCP1 (43.13 ± 1.35 D), TCP2 (41.87 ± 1.30 D), and TCP-IOL (42.62 ± 1.35 D) were significantly different (P < .0001). The best results were obtained using simulated K: the median absolute error ranged between 0.22 D and 0.29 D and the percentage of eyes with a prediction error of ±0.50 D or less, between 76.2% and 84.7%, depending on the formula. After constant optimization, the results using any TCP value and simulated K were similar with no statistically significant differences.
Conclusions:
Biometric measurements provided by the Scheimpflug-Placido optical biometer can be used to accurately calculate the IOL power. Simulated K and TCP led to similar outcomes after constant optimization.
More Related Videos
Related Concept Videos
Power
Sums of Power
Instantaneous Power
Complex Power
Complex power is defined as the multiplication of the voltage and the complex conjugate of the current. The magnitude of this power, known as apparent power, is measured in volt-amperes (VA). Notably, the angle of the complex power equates to the...
Electrical Power
Nuclear Power
Nuclear Fuels
Nuclear fuel consists of a fissile isotope, such as uranium-235, which must be present in sufficient quantity to provide a...