Related Experiment Video
Updated: Feb 9, 2026

A Human Ex Vivo Atherosclerotic Plaque Model to Study Lesion Biology
Published on: May 6, 2014
[Study of New Micropore RF system on Lesion Formation and Complications]
Yuwen Song1, Xiulin Xu1, Yameng Cai2
1University of Shanghai for Science and Technology, Shanghai, 200093.
Objectives:
To study the safety and effectiveness of a new type of micropore ablation catheter in vitro ablation system, and to provide reference for clinical practice.
Methods:
To evaluate two kinds of catheter in cardiac tissue ablation depth, tissue temperature and thrombosis situation by the same RF system.
Results:
The power set 25 W, There was no significant difference in ablation depth between the two groups, and no Pop and thrombosis occurred. When the power is more than 40 W, two groups occurred more Pop and thrombosis.
Conclusions:
When using high power for Cardiac RF ablation, doctors should pay more attention to complications and thrombosis.