Purpose:
To investigate the dosimetric impact of hysteresis on lung cancer tomotherapy.
Methods:
Measurements were acquired using MapCheck with an XY4D motion simulation table. Six hysteresis states (0, π/32, π/16, π/8, 3π/16, and π/4) were considered with sinusoidal motions in the superior-inferior and left-right orientations. The measured data were analyzed both globally (from all detectors) and structure-by-structure in the measurement plane. The dose difference (DD) analysis method with local normalization in the absolute dose mode with a DD threshold of 6 cGy was adopted to analyze each hysteresis vs. static state (H(∗)S) and nonzero vs. zero hysteresis (H(∗)0). The threshold was 10% for all analyses. Wilcoxon signed rank tests with significance level p = 0.05 were used for statistical analysis.
Results:
The DD analysis of each H(∗)S mostly indicated that the passing rate differed between structures but was similar between hysteresis states. The DD analysis of H(∗)0 showed that the passing rate decreased with increasing hysteresis. The differences between larger hysteresis (≥3π/16) and other states were significant for comparisons between global, left lung, chest wall, and target. Both analyses showed that the DD distribution changed with hysteresis.
Conclusions:
Hysteresis difference causes the DD distribution to change. Structural difference had more impact than hysteresis state difference on hysteresis motion vs. static comparisons. Remarkable effects on nonzero vs. zero hysteresis comparisons were only seen for structures closely related to the target at large hysteresis. Small organs at risk that are close to the target need to be considered further.
More Related Videos
08:30Intraperitoneal Glucose Tolerance Test, Measurement of Lung Function, and Fixation of the Lung to Study the Impact of Obesity and Impaired Metabolism on Pulmonary Outcomes
Published on: March 15, 2018
08:34Visualization of High Speed Liquid Jet Impaction on a Moving Surface
Published on: April 17, 2015
Related Concept Videos
Lung Capacity
Impact of Groups on Groups
Mouse Models of Cancer Study
The development of transgenic, knockout, and knock-in mice has led to an exponential increase in their use as model organisms in research,...
Impact
When particles with different initial velocities collide, they induce deformation by applying equal and opposite impulses. At the point of maximum deformation, the particles move together with...
Impact of Groups on Individuals
Impact of Individuals on a Group