A Protocol to Set Up Needle-Free Connector with Positive Displacement on Central Venous Catheter in Intensive Care Unit
Published on: July 13, 2019
Variability in alignment of central venous pressure transducer to physiologic reference point in the intensive care
Carl Sjödin1, Soren Sondergaard2, Lotta Johansson3
1Sahlgrenska University Hospital, Gothenburg, Sweden.
Background:
The phlebostatic axis is the most commonly used anatomical external reference point for central venous pressure measurements. Deviation in the central venous pressure transducer alignment from the phlebostatic axis causes inadequate pressure readings, which may affect treatment decisions for critically ill patients in intensive care units.
Aim:
The primary aim of the study was to assess the variability in central venous pressure transducer levelling in the intensive care unit. We also assessed whether patient characteristics impacted on central venous pressure transducer alignment deviation.
Methods:
A sample of 61 critical care nurses was recruited and asked to place a transducer at the appropriate level for central venous pressure measurement. The measurements were performed in the intensive care unit on critically ill patients in supine and Fowler's positions. The variability among the participants using eyeball levelling and a laser levelling device was calculated in both sessions and adjusted for patient characteristics.
Results:
A significant variation was found among critical care nurses in the horizontal levelling of the pressure transducer placement when measuring central venous pressure in the intensive care unit. Using a laser levelling device did not reduce the deviation from the phlebostatic axis. Patient characteristics had little impact on the deviation in the measurements.
Conclusion:
The anatomical external landmark for the phlebostatic axis varied between critical care nurses, as the variation in the central venous pressure transducer placement was not reduced with a laser levelling device. Standardisation of a zero-level for vascular pressures should be considered to reduce the variability in vascular pressure readings in the intensive care unit to improve patient treatment decisions. Further studies are needed to evaluate critical care nurses' knowledge and use of central venous pressure monitoring and whether assistive tools and/or routines can improve the accuracy in vascular pressure measurements in intensive care units.
Related Concept Videos
Correlations
Venous Thrombosis III: Interprofessional Care
Intensity and Pressure of Sound Waves
Unlike the time average of a sinusoidal term, which is zero since it is positive...
Fluid Pressure over Flat Plate of Variable Width
The pressure distribution on the plate can be calculated by determining the force that acts on a differential area strip of the plate. Thus, the magnitude of the force is equal to the...
Correlation and Causation
Correlation versus Causation
If the dependent variable increases or decreases when the independent variable increases, there is a positive or negative...
Measurement: Derived Units