Related Experiment Video
Updated: Feb 9, 2026

Evaluation of Left Ventricular Structure and Function using 3D Echocardiography
Published on: October 28, 2020
Evaluation of right ventricular function by coronary computed tomography angiography using a novel automated 3D right
Philipp Burghard1,2, Fabian Plank3, Christoph Beyer4
1Department of Radiology, Innsbruck Medical University, Anichstrasse 35, 6020, Innsbruck, Austria. philipp.burghard@hotmail.com.
Objectives:
To evaluate right ventricle (RV) function by coronary computed tomography angiography (CTA) using a novel automated three-dimensional (3D) RV volume segmentation tool in comparison with clinical reference modalities.
Methods:
Twenty-six patients with severe end-stage heart failure [left ventricle (LV) ejection fraction (EF) <35%] referred to CTA were enrolled. A specific individually tailored biphasic contrast agent injection protocol was designed (80%/20% high/low flow) was designed. Measurement of RV function [EF, end-diastolic volume (EDV), end-systolic volume (ESV)] by CTA was compared with tricuspid annular plane systolic excursion (TAPSE) by transthoracic echocardiography (TTE) and right heart invasive catheterisation (IC).
Results:
Automated 3D RV volume segmentation was successful in 26 (100%) patients. Read-out time was 3 min 33 s (range, 1 min 50s-4 min 33s). RV EF by CTA was stronger correlated with right atrial pressure (RAP) by IC (r = -0.595; p = 0.006) but weaker with TAPSE (r = 0.366, p = 0.94). When comparing TAPSE with RAP by IC (r = -0.317, p = 0.231), a weak-to-moderate non-significant inverse correlation was found. Interobserver correlation was high with r = 0.96 (p < 0.001), r = 0.86 (p < 0.001) and r = 0.72 (p = 0.001) for RV EDV, ESV and EF, respectively. CT attenuation of the right atrium (RA) and right ventricle (RV) was 196.9 ± 75.3 and 217.5 ± 76.1 HU, respectively.
Conclusions:
Measurement of RV function by CTA using a novel 3D volumetric segmentation tool is fast and reliable by applying a dedicated biphasic injection protocol. The RV EF from CTA is a closer surrogate of RAP than TAPSE by TTE.
Key Points:
• Evaluation of RV function by cardiac CTA by using a novel 3D volume segmentation tool is fast and reliable. • A biphasic contrast agent injection protocol ensures homogenous RV contrast attenuation. • Cardiac CT is a valuable alternative modality to CMR for the evaluation of RV function.
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Reliability and Validity
Acute Coronary Syndrome III: Diagnostic Studies
Anatomy of the Brain: Ventricles
Coronary Circulation
Coronary circulation begins at the base of the aorta, where two main arteries arise—the left and right coronary arteries. These arteries encircle the heart in the coronary sulcus and supply the...