A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Change of sensing vector in the subcutaneous ICD during follow-up and after device replacement
Markus Bettin1, Benjamin Rath1, Christian Ellermann1
1Department of Cardiology II - Electrophysiology, University Hospital Münster, Münster, Germany.
Background:
The subcutaneous implantable cardioverter defibrillator (S-ICD) has been established as a valuable alternative to transvenous ICD for prevention of sudden cardiac death. The system automatically chooses the optimal sensing vector. However, during follow-up and especially after device replacement we observed a change of the suggested sensing vector in automatic setup. Therefore, we analyzed frequency and reasons of vector change and its impact on inappropriate shocks (IAS).
Material And Methods:
Between June 2010 and December 2017, a total of 216 patients with S-ICD® were included in this analysis. In all patients sensing vectors at the time of implantation, during follow-up, and after device replacement were investigated. Median follow-up time was 27.3 ± 25.3 months.
Results:
A change of the initial vector was seen in 77 patients (35.7%). The most frequent reason for vector change was the postoperative setup in supine and erect position in 54 patients (70.1%). In 12 patients (15.5%), the vector was manually changed due to inappropriate sensing and/or therapies. Routine setup during follow-up led to automatic vector change in 10 cases (13.0%). In only 1 patient the vector was manually changed due to oversensing in an exercise treadmill test. In 27 patients, the device was replaced due to battery depletion and in 6 of these patients the sensing vector was changed by the automatic setup. Vector change did not have an impact for inappropriate therapies in the follow-up; only 1 patient received an IAS due to an inadvertent vector change after device replacement.
Conclusion:
In the present study, a significant number of S-ICD® patients had a manual or automatic vector change during follow-up and after device replacement. The study underlines the importance of a thoroughly performed screening and at least two valuable sensing vectors preimplant. Further studies are needed to evaluate the necessity of a routine automatic setup during follow-up.
Related Concept Videos
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Introduction to Special Senses
Global Climate Change
Rates of Change
Work Done During Volume Change
Consider a gas confined to a cylinder fitted with a movable piston at one end. If the gas expands from volume V1 to volume V2, it exerts a force on the piston, such that the piston moves by a distance dr.
The work done by the gas on the piston can be expressed as
Le Chatelier's Principle: Changing Temperature
To understand this phenomenon, consider the elementary reaction: