Calibration Procedures for Orthogonal Superposition Rheology
Published on: November 18, 2020
Calibration and validation of Airborne Multiangle SpectroPolarimetric Imager (AirMSPI) polarization measurements
Abstract:
The Airborne Multiangle SpectroPolarimetric Imager (AirMSPI), a precursor to the future Multi-Angle Imager for Aerosols satellite instrument, is a remote-sensing instrument for the characterization of atmospheric aerosols and clouds. To help discriminate between different aerosol particle types, which is crucial to improve our understanding of their impact on climate and air quality, AirMSPI acquires imagery over multiple view angles in the ultraviolet, visible, and near-infrared, and it employs dual photoelastic modulators (PEMs) to target an uncertainty requirement of ±0.005 in the degree of linear polarization (DoLP) at selected wavelengths. Laboratory polarimetric calibrations using a second-generation Polarization State Generator-2 (PSG-2) and validation measurements at 0