Objectives:
Several procedures have been described to correct pectus excavatum and carinatum. We have used a modified Ravitch procedure (STRATOS titanium bars) for patients who were unsuitable for the minimally invasive Nuss procedure. The operation produced excellent cosmetic results, but we have noted several fractures and displacements of the STRATOS bars.
Methods:
We reviewed all our STRATOS patients since first use in 2009 until 2014. We collated the following data of each patient: diagnosis, severity of pectus deformity, comorbidity, previous operations for pectus, age at bar implantation, the number of bars implanted and duration of implantation. Patients with severe comorbidity were excluded.
Results:
Between 2009 and 2014, 39 patients were treated for pectus deformities using the STRATOS bars. Twelve of 39 patients (31%) were identified as having bar(s) broken before removal. There were no significant differences between patients with fractured bars and those without, except for the number of bars implanted (P = 0.016).
Conclusions:
This incidence of bar fracture is unexpectedly high, although the literature on this topic is sparse. The high fracture rate is alarming, because of its unpredictability and lack of clear aetiological factors. The operation is done (largely) for psychosocial and cosmetic reasons, and therefore, we must have complete confidence in the quality of the bars used in teenage pectus patients. We raise the issue to warn other units and encourage them to report bar fracture rates.
More Related Videos
15:11Surgical Fixation of Sternal Fractures: Preoperative Planning and a Safe Surgical Technique Using Locked Titanium Plates and Depth Limited Drilling
Published on: January 5, 2015
06:59Fracture Apparatus Design and Protocol Optimization for Closed-stabilized Fractures in Rodents
Published on: August 14, 2018
Related Concept Videos
Ecological Disturbance
Bar Graph
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Multiple Bar Graph
Each bar or column in the multiple bar graph represents a data value. These graphs are used primarily in interrelating two or more sets of data. The categories of different kinds of data are listed along the horizontal or x-axis, whereas...
Distance Corrections
Power Factor Correction