Quantitative ultrasound imaging over the ischial tuberosity: An exploratory study to inform tissue health
Sharon Gabison1, Sunita Mathur2, Mary C Verrier1
1Department of Physical Therapy, University of Toronto, 500 University Avenue, Toronto, Ontario, M5G 1V7, Canada; SCI Mobility Lab, Lyndhurst Centre, Toronto Rehab-University Health Network, 520 Sutherland Drive, Room 236, Toronto, Ontario, M4G 3V9, Canada.
Background:
Characterization of ischial tissue health using a standardized diagnostic ultrasound protocol capturing thickness and gray scale analysis has not been established.
Objectives:
This study evaluates inter-participant and inter-trial reliability of thickness and gray scale analysis of ultrasound images of tissues overlying the ischial tuberosity. It provides recommendations for the number of images required to minimize the standard error of measurement (SEM) and determines the number of images required for thickness, gray scale and contrast values that exceed an a-priori minimal detectable change (MDC) for repeated tissue assessment.
Methods:
Brightness mode ultrasound images using a 12 MHz linear probe were collected on the dominant limb in the side lying position for ten healthy participants and partitioned into three regions of interest: skin, subcutaneous tissue and muscle. Thickness and gray scale measures of skin, muscle and subcutaneous tissue were calculated using a customized MATLAB program. Contrast of each region of interest was calculated using the Gray Scale Level Co-Occurrence Matrix. Generalizability theory was used to quantify indices of dependability and corresponding SEMs and MDCs with 90% Confidence Intervals.
Results:
Participants accounted for most of the total variance (75.56% to 94.78%). Coefficient of dependability (ϕ) for thickness, grey scale and contrast measures was greater than 0.80 when more than two images were averaged. In order to detect a MDC of 21% in thickness and echogenicity measures, at least three images are required, while at least 5 images are required for a MDC of 25% for contrast measures.
Conclusions:
Obtaining reliable thickness, echogenicity and contrast measures of tissue overlying the ischial tuberosity can be achieved from two ultrasound images by a single therapist on an individual participant however three and five images are required to use a MDC of 21% for thickness measures and MDC of 25% for contrast measures respectively.
Related Concept Videos
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Health Literacy
Imaging Studies VII: Vascular Imaging
Imaging Studies II: Ultrasonography
Imaging Studies I: Kidney, Ureter, and Bladder Studies
Purpose of Health Records I
Here's a breakdown of how health records serve these purposes: