Validation of a Simple Overlay Device to Assess Radial Head Implant Length
Dave R Shukla1, Matthias K D Vanhees1, James S Fitzsimmons1
1Biomechanics Laboratory, Department of Orthopedic Surgery, Mayo Clinic, Rochester, MN.
Purpose:
A simple overlay device (SOD) was developed to measure radial head implant length. The purpose of this study was to determine the accuracy and reliability of this device for measuring experimental radial head implant length.
Methods:
Five fresh frozen cadavers were implanted with sequentially longer implants, adjusted by neck length (0, 2, 4, and 8 mm). Fluoroscopic images were obtained in 4 forearm positions: anteroposterior in supination in full extension, anteroposterior in pronation in full extension, supinated in 45° of flexion, and neutral in 45° of flexion. The SOD measurements (made by 2 observers) were compared with the native original radial head (control) to assess implant length. In addition, gapping of the ulnohumeral joint space was measured for comparison purposes.
Results:
The measured radial head and neck lengths for the specimens were 33, 39, 31, 34, and 42 mm. The difference between the actual radial head and neck lengths and those measured with the SOD template averaged less than 2 mm for all 4 collar sizes, except in 1 measurement in which the bicipital tuberosity could not be visualized. The median intraclass correlation coefficients for observer 1 compared with the SOD were 0.94 to 0.99. The median intraclass correlation coefficients between observers were 0.88 to 0.95. For both observers, elbow position, collar height, and the 2 variables combined did not significantly affect the SOD values. The other method that was evaluated, that of measurement of the ulnohumeral joint space, had higher interobserver variability versus the SOD, and allowed detection of lengthening of over 4 mm.
Conclusions:
The SOD is a reliable method for simply assessing radial head length with radiographs and can accurately detect 2 mm or more of proximal radial lengthening.
Clinical Relevance:
The SOD is a simple and accurate method that can help to optimize radial head sizing.
More Related Videos
04:35Author Spotlight: Innovative Methodology for Implanting and Securing Neural Probes in the Rodent Spinal Cord
Published on: July 12, 2024
08:07Author Spotlight: Advancing Traumatic Brain Injury Research - A Closed-Head Model for Accurate Replication and Rapid Assessment
Published on: September 22, 2023
Related Concept Videos
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps:
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Reliability and Validity
Radial System Protection
In a radial system with a fault downstream of the third breaker, ideally, only the third breaker will open, isolating the fault and interrupting the load connected beyond it. The second breaker has a longer delay setting,...
Muscles that Move the Head
The bilateral sternocleidomastoid, or SCM, and the suprahyoid and infrahyoid muscles are significant head flexors. The SCM muscles originate at the sternum and clavicle and attach to the mastoid process of the temporal bone. The SCM contracts bilaterally to bend the head forward, whereas...
Data Validation
Key parameters for method validation include: