Effect of defibrillation threshold testing on effectiveness of the subcutaneous implantable cardioverter
Lakshmi Peddareddy1, Faisal M Merchant2, Angel R Leon2
1Emory University School of Medicine, Atlanta, GA, USA.
Background:
Defibrillation threshold (DFT) testing is recommended with the subcutaneous implantable cardioverter defibrillator (SICD).
Objective:
To describe first shock efficacy for appropriate SICD therapies stratified by the presence of implant DFT testing.
Methods:
We reviewed all patients receiving SICDs at our institution and stratified them based on whether implant DFT testing was performed. Appropriate shocks were reviewed to see if ventricular tachycardia/ventricular fibrillation (VT/VF) terminated with a single shock. First shock efficacy was stratified by implant DFT status.
Results:
178 patients implanted with SICDs and followed in our center were included in this study. Of these, 135 (76%) underwent DFT testing (DFT (+) group). In the DFT (+), 80 appropriate shocks were needed to treat 69 episodes of VT/VF. The first shock was effective in 61 out of 69 episodes (88.4%), whereas multiple shocks were required to terminate VT/VF in the remaining eight episodes. Among 43 patients without implant DFT testing (DFT (-) group), 20 appropriate shocks to treat 17 episodes of VT/VF occurred in seven patients. VT/VF was successfully terminated with the first shock in 16 out of 17 episodes (first shock efficacy 94.1 %). There was no significant difference in first shock effectiveness between those with and without implant DFT testing (P = 0.97).
Conclusion:
A strategy that omits DFT testing at implant did not appear to compromise the effectiveness of the SICD. These data suggest that routine DFT testing at SICD implant might not be necessary. Randomized trials are needed to confirm this finding.
Related Concept Videos
Buffer Effectiveness
The buffer capacity is the amount of acid or base that can be added to a given volume...
Framing Effects
Biological Effects of Radiation
Testes: Histology
The spermatogenic cells, responsible for producing sperm, are...
Complementation Tests
Organisms heterozygous for different mutations are crossed pairwise in all combinations. If present on different genes, the mutations can complement each other by providing the missing...
Effects of feedback
Feedback significantly modifies the gain of a control system. The gain of a system without feedback is altered by a factor of one plus GH, where G represents...