A Comparison of the Diagnostic Accuracy of Common Office Blood Pressure Measurement Protocols
Ian M Kronish1, Donald Edmondson1, Daichi Shimbo1
1Center for Behavioral Cardiovascular Health, Department of Medicine, Columbia University Medical Center, New York, New York, USA.
Background:
The optimal approach to measuring office blood pressure (BP) is uncertain. We aimed to compare BP measurement protocols that differed based on numbers of readings within and between visits and by assessment method.
Methods:
We enrolled a sample of 707 employees without known hypertension or cardiovascular disease, and obtained 6 standardized BP readings during each of 3 office visits at least 1 week apart, using mercury sphygmomanometer and BpTRU oscillometric devices (18 readings per participant) for a total of 12,645 readings. We used confirmatory factor analysis to develop a model estimating "true" office BP that could be used to compare the probability of correctly classifying participants' office BP status using differing numbers and types of office BP readings.
Results:
Averaging 2 systolic BP readings across 2 visits correctly classified participants as having BP below or above the 140 mm Hg threshold at least 95% of the time if the averaged reading was <134 or >149 mm Hg, respectively. Our model demonstrated that more confidence was gained by increasing the number of visits with readings than by increasing the number of readings within a visit. No clinically significant confidence was gained by dropping the first reading vs. averaging all readings, nor by measuring with a manual mercury device vs. with an automated oscillometric device.
Conclusions:
Averaging 2 BP readings across 2 office visits appeared to best balance increased confidence in office BP status with efficiency of BP measurement, though the preferred measurement strategy may vary with the clinical context.
Related Concept Videos
Measurement of Blood Pressure
Uncertainty in Measurement: Accuracy and Precision
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Blood Pressure
Blood Pressure
The average BP in an adult is typically around 120/80 mmHg (millimeters of mercury). In this measurement, the numerator (120) indicates the systolic pressure, which is the pressure in the arteries during the contraction of the heart's ventricles as blood is expelled. The denominator (80) represents the...