Verification study on the NORIP LDH reference intervals with a proposed new upper reference limit.
Søren Kristiansen1, Lennart Friis-Hansen1, Claus Antonio Juel Jensen1
1a Department of Clinical Biochemistry , Copenhagen University Hospital of Nordsjælland , Denmark.
This study verified the Nordic Reference Interval Project (NORIP) lactate dehydrogenase (LDH) reference interval. Findings suggest the NORIP upper limit for LDH may need adjustment due to a high percentage of results exceeding it.
Area of Science:
- Clinical Chemistry
- Laboratory Medicine
- Biochemistry
Background:
- The Nordic Reference Interval Project (NORIP) provides reference intervals (RI) for laboratory tests.
- Lactate dehydrogenase (LDH) is an enzyme found in many body tissues, and its levels can indicate cellular damage or disease.
Purpose of the Study:
- To verify the established serum lactate dehydrogenase (LDH) reference interval (RI) from the Nordic Reference Interval Project (NORIP).
- To assess the clinical applicability of the NORIP RI for LDH by comparing it with locally derived data.
Main Methods:
- Serum LDH concentrations were analyzed using the Dimension Vista 1500 system with an IFCC method.
- Retrospective data mining of clinical laboratory information systems was employed to calculate an indirect RI for LDH.
- Comparison of the calculated indirect RI with the existing NORIP RI for the adult age group (18 to <70 years).
Main Results:
- The study verified the transference of the NORIP RI for LDH, with minor bias observed against reference serums.
- An indirectly calculated RI for adults showed an upper limit of 240 U/L (90% CI: 234-243 U/L).
- A significant percentage (25.1%) of retrospective LDH test results exceeded the NORIP upper limit of 205 U/L, with variations based on the requesting clinical setting.
Conclusions:
- The current NORIP upper limit for serum LDH (205 U/L) may be too low, as evidenced by a high proportion of results exceeding this threshold.
- The findings suggest a need to re-evaluate and potentially adjust the NORIP upper RI for LDH to better reflect current clinical laboratory data.
- Further investigation into factors influencing LDH levels and reporting is warranted.
Related Concept Videos
Inertial Frames of Reference
Non-inertial Frames of Reference
Chemical Shift: Internal References and Solvent Effects
The internal reference compound generally used in NMR spectroscopy is tetramethylsilane (TMS). TMS is preferred because it is chemically inert, soluble in NMR solvents, and easily removable. Also, the highly shielded methyl protons in TMS yield an intense...
Self-Evaluation: Self-Enhancement and Self-Verification
Prediction Intervals
However, the point estimate is most likely not the exact value of the population parameter, but close to it. After calculating point estimates, we construct interval estimates, called confidence intervals or prediction intervals. This prediction interval comprises a range of values unlike the point estimate and is a better predictor of the observed sample value, y.
Strategies of Self-Presentation II: Self-Verification