Related Experiment Video
Updated: Feb 9, 2026

Fully Automated Centrifugal Microfluidic Device for Ultrasensitive Protein Detection from Whole Blood
Published on: April 16, 2016
Antecedent rest may not be necessary for automated office blood pressure at lower treatment targets
Tracey J F Colella1,2,3, Anam Tahsinul1, Hannah Gatto1
1Cardiovascular Prevention and Rehabilitation Program, University Health Network/Toronto Rehabilitation Institute, Toronto, ON, Canada.
Abstract:
In SPRINT (Systolic Blood Pressure Intervention Trial), use of the Omron 907XL blood pressure (BP) monitor set at 5 minutes of antecedent rest to record BP produced an automated office BP value 7/6 mm Hg lower than awake ambulatory BP at 27 months. The authors studied the impact on automated office BP of setting the Omron 907XL to 0 minutes instead of 5 minutes of rest in patients with readings in the lower normal BP range, similar to on-treatment BP in the SPRINT intensive therapy group. Patients (n = 100) in cardiac rehabilitation were randomized to three BP readings at 1-minute intervals using an Omron 907XL BP device set for 5 or 0 minutes of antecedent rest. Mean (±standard deviation) automated office BP (mm Hg) after 5 minutes of rest (120.2 ± 14.6/66.9 ± 8.6 mm Hg) was lower (P < .001/P < .01) than without rest (124.2 ± 16.4/67.9 ± 9.1 mm Hg). When target BP is in the lower normal range, automated office BP recorded without antecedent rest using an Omron 907XL device should be higher and closer to the awake ambulatory BP, compared with readings taken after 5 minutes of rest.
Related Concept Videos
Pressure Variation in a Fluid at Rest
When measuring pressure at two different levels within the fluid, the difference in...
Blood Pressure
Blood Pressure
The average BP in an adult is typically around 120/80 mmHg (millimeters of mercury). In this measurement, the numerator (120) indicates the systolic pressure, which is the pressure in the arteries during the contraction of the heart's ventricles as blood is expelled. The denominator (80) represents the...
Measurement of Blood Pressure
Factors affecting Blood pressure
Physiological Factors:
Alterations in Blood Pressure
Hypertension (High blood pressure)