[Method for continuous measurement of personal exposure to fine particulate matter]
Na Li1, Zhe Liu1, Yibing Yang1
1Institute for Environmental Health and Related Product Safety, Chinese Center for Disease Control and Prevention, Beijing 100021, China.
Objective:
To develop a new accurate method for continuous measurement of personal exposure to fine particulate matter( PM_(2. 5) ) by integrating the advantage of both light scattering and gravimetric method.
Methods:
The PM_(2. 5) concentrations were measured by both light scattering and gravimetric method simultaneously. The ratio of time-weighted average concentrations from gravimetric method to the responding values from light scattering method was calculated as coefficient of correction and used to calibrate the real-time PM_(2. 5) concentrations from light scattering method.
Results:
The limit of detections( LOD) for 3, 10, 30 and 60 min time-weighted average concentrations of PM_(2. 5) were 4. 6, 4. 0, 3. 9 and 3. 7 μg/m~3, respectively. The precision( relativestandard deviation, RSD) was from 2. 1% to 9. 5%. The 30 min time-weighted average concentrations from this method were highly related to the results from TEOM method which was considered as reference method( Pearson r = 0. 934, P < 0. 001, n = 233), and no significantly bias between these two methods was detected by paired t test( P =0. 957). Zero drift was detected for 3 out of 10 devices with drift values in the range of- 5-- 3 μg/m~3 in the experiment of continuous 24 h monitoring conducted in the lab. And zero shift was also investigated in 7. 5%( 31 /412) field monitoring events, but most of the values shifted were within the range of- 3- 3 μg/m~3. The direction and scale of zero shift were not influenced by PM_(2. 5) concentrations.
Conclusion:
The developed method in this study that combines the advantages of both light-scattering and gravimetric method is validated to measure real-time concentrations of personal exposure to PM_(2. 5) .
Related Concept Videos
Classifying Matter by State
Physical and Chemical Properties of Matter
Classifying Matter by Composition
According to its composition, the matter can be classified into two broad categories — pure substances and mixtures.
A pure substance is a form of matter that has a constant composition throughout with uniform properties. For example, any sample of sucrose has the same composition and same physical properties, such as melting point, color, and sweetness, regardless of the source from which it is isolated.
A mixture is composed of two or...
What is Matter?
The Atomic Theory of Matter
Fineness of Cement
Direct...