Evaluation and verification of the QFix EncompassTM couch insert for intracranial stereotactic radiosurgery
Karen Chin Snyder1, Ilma Xhaferllari1, Yimei Huang1
1Department of Radiation Oncology, Henry Ford Health System, Detroit, MI, USA.
Abstract:
The QFix EncompassTM stereotactic radiosurgery (SRS) immobilization system consists of a thermoplastic mask that attaches to the couch insert to immobilize patients treated with intracranial SRS. This study evaluates the dosimetric impact and verifies a vendor provided treatment planning system (TPS) model in the Eclipse TPS. A thermoplastic mask was constructed for a Lucy 3D phantom, and was scanned with and without the EncompassTM system. Attenuation measurements were performed in the Lucy phantom with and without the insert using a pinpoint ion chamber for energies of 6xFFF, 10xFFF and 6X, with three field sizes (2 × 2, 4 × 4, and 6 × 6 cm2 ). The measurements were compared to two sets of calculations. The first set utilized the vendor provided Encompass TPS model (EncompassTPS ), which consists of two structures: the Encompass and Encompass base structure. Three HU values for the Encompass (200, 300, 400) and Encompass Base (-600, -500, -400) structures were evaluated. The second set of calculations consists of the Encompass insert included in the external body contour (EncompassEXT ) for dose calculation. The average measured percent attenuation in the posterior region of the insert ranged from 3.4%-3.8% for the 6xFFF beam, 2.9%-3.4% for the 10xFFF, and 3.3%-3.6% for the 6X beam. The maximum attenuation occurred at the region where the mask attaches to the insert, where attenuation up to 17% was measured for a 6xFFF beam. The difference between measured and calculated attenuation with either the EncompassEXT or EncompassTPS approach was within 0.5%. HU values in the EncompassTPS model that provided the best agreement with measurement was 400 for the Encompass structure and -400 for the Encompass base structure. Significant attenuation was observed at the area where the mask attaches to the insert. Larger differences can be observed when using few static beams compared to rotational treatment techniques.
More Related Videos
10:52Stereotactic Intracranial Implantation and In vivo Bioluminescent Imaging of Tumor Xenografts in a Mouse Model System of Glioblastoma Multiforme
Published on: September 25, 2012
06:44Modeling Brain Metastases Through Intracranial Injection and Magnetic Resonance Imaging
Published on: June 7, 2020
Related Concept Videos
Self-Evaluation: Self-Enhancement and Self-Verification
Strategies of Self-Presentation II: Self-Verification
Porin Insertion in the Outer Mitochondrial Membrane
Three models describe the assembly of porins by the SAM complex and their insertion into the outer membrane. Model 1 suggests that porins are assembled outside the SAM channel as the...
Insertion of Single-pass Transmembrane Proteins in the RER
Integral transmembrane proteins possess transmembrane and extra membrane domains. The transmembrane domains are primarily made of 20-25 hydrophobic amino acids arranged in a helical secondary confirmation. These...
Insertion of Multi-pass Transmembrane Proteins in the RER
The multipass transmembrane proteins are the type IV integral membrane proteins with multiple topogenic sequences determining their spatial arrangement in the ER membrane. Nearly all multipass proteins lack a cleavable signal sequence and use...
Nursing Evaluation