Related Experiment Video
Updated: Feb 8, 2026

Echocardiographic Evaluation of Atrial Communications before Transcatheter Closure
Published on: February 8, 2022
Comprehensive Echocardiographic Assessment of Normal Transcatheter Valve Function
Rebecca T Hahn1, Jonathon Leipsic2, Pamela S Douglas3
1Columbia University Medical Center/New York-Presbyterian Hospital, New York, New York.
This study establishes normal hemodynamic parameters for balloon-expandable and self-expanding transcatheter aortic valves. These findings aid in assessing valve function and guiding clinical decisions for improved patient outcomes.
Area of Science:
- Cardiovascular Medicine
- Biomedical Engineering
- Interventional Cardiology
Background:
- Transcatheter aortic valve replacement (TAVR) requires reliable methods to assess prosthetic valve function.
- Echocardiographic Doppler-derived hemodynamic data are crucial for pre- and post-implant evaluation of TAVR devices.
- Establishing normal hemodynamic parameters is essential for identifying structural valve dysfunction.
Purpose of the Study:
- To define normal hemodynamic function parameters for different iterations of balloon-expandable and self-expanding transcatheter aortic valves.
- To provide expected echocardiographic Doppler-derived hemodynamic data for TAVR devices.
- To inform pre-implant decision-making and post-implant monitoring of longitudinal valve function.
Main Methods:
- Collected echocardiography core lab data on mean gradients and effective orifice area (EOA) from discharge or 30-day echocardiograms.
- Analyzed data from randomized trials including PARTNER (balloon-expandable) and Medtronic CoreValve US Pivotal/Evolut R IDE studies (self-expanding).
- Examined relationships between valve size, annular anatomy, and post-implant EOA.
Main Results:
- Provided mean EOA and gradient values for SAPIEN, SAPIEN XT, SAPIEN 3, CoreValve, and Evolut R valves.
- SAPIEN 3 EOA increased with baseline annular area (p < 0.001).
- Evolut R EOA increased with baseline annular perimeter (p < 0.001).
Conclusions:
- Tables of expected transcatheter aortic valve hemodynamics by type and size are vital for function evaluation.
- Expected EOA data correlated with native annular anatomy, aiding pre-implantation decisions.
- Proposed criteria for defining structural valve dysfunction based on these hemodynamic parameters.
Related Concept Videos
Mitral Valve Prolapse II: Assessment and Management
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Normal Distribution
Normal Stress
When a rod is under axial loading, the internal forces and corresponding stress are normal to the plane of the section, so it is termed normal stress. It's important to...
Applications of Normal Distribution
The heights of 15 to 18-year-old males from Chile from 1984 to 1985 followed a normal distribution. The mean height is 172.36...
Introduction to Normal Distributions