Semiquantitative analysis using standardized uptake value in 123I-FP-CIT SPECT/CT
Related Experiment Video
Updated: Feb 8, 2026

A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Akira Toriihara1, Hiromitsu Daisaki2, Akihiro Yamaguchi3
1Department of Diagnostic Radiology and Nuclear Medicine, Tokyo Medical and Dental University, 1-5-45, Yushima, Bunkyo-ku, Tokyo 113-8519, Japan.
Purpose:
To evaluate potential of a semiquantitative method using standardized uptake value (SUV) in 123I-2β-carbomethoxy-3β-(4-iodophenyl)-N-(3-fluoropropyl) nortropane (123I-FP-CIT) single photon emission computed tomography/computed tomography (SPECT/CT) compared with specific binding ratio (SBR).
Materials And Methods:
First, we performed a phantom study to validate the accuracy of measuring SUV. 52 patients (25 male, 27 female; mean age of 75.1-year-old; 40 and 12 patients with neurodegenerative diseases with or without presynaptic dopaminergic deficits, respectively) were enrolled in a retrospective study. We measured SBR, maximum SUV, peak SUV, mean SUV, and striatum-to-background ratio of SUV (SUVratio) for striatum with lower 123I-FP-CIT uptake using commercial software. We calculated Pearson's correlation coefficient between SBR and SUV. We also calculated the sensitivity, specificity, and accuracy of each parameter for differential diagnosis.
Results:
The phantom study revealed errors of <10% between theoretical and actual SUVs. Although there were significant correlations between SBR and all SUV-based parameters, SUVratio showed the most strong correlation with SBR (r = 0.877, p < 0.001). However, diagnostic capability of SUVratio (cutoff = 2.35) yielded to that of SBR (cutoff = 3.90) for diagnosing neurodegenerative diseases with presynaptic dopaminergic deficits (sensitivity of 85.0% vs 92.5%, specificity of 100% vs 91.7%, and accuracy of 88.5% vs 92.3%, respectively).
Conclusion:
SBR is a promising parameter to aid differential diagnosis of neurodegenerative diseases with or without presynaptic dopaminergic deficit. Although technically acceptable, SUV may not be superior to SBR when clinically applied in 123I-FP-CIT SPECT/CT.
Related Concept Videos
Standard Enthalpy of Formation
Standard Electrode Potentials
Measurement: Standard Units
Radiological Investigation I: X-ray and CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Calculating Standard Free Energy Changes