Influence of acquisition parameters on the magnitude of cone beam computed tomography artifacts
Deborah Queiroz Freitas1, Rocharles Cavalcante Fontenele1, Eduarda Helena Leandro Nascimento
11 Department of Oral Diagnosis, Division of Oral Radiology, Piracicaba Dental School, University of Campinas , Limeira , Brazil.
Aim:
To evaluate the influence of kilovoltage (kVp) and metal artifact reduction tool (MAR) on the magnitude of cone beam CT (CBCT) artifacts.
Methods:
A titanium and zirconia implants were inserted alternately in a posterior region of a mandible. CBCT exams were acquired with ProMax 3D (Planmeca Oy, Helsinki, Finland) and Picasso Trio machines (Vatech, Hwaseong, South Korea) using 70 kVp, 80 kVp and 90 kVp with and without MAR activation. The other exposure factors remained fixed at 5mA, field of view 80 × 50 mm and voxel 0.20 mm. The scans were performed before and after the insertion of the implants. Regions of interest were determined in different distances from the artifact production area (15, 25 and 35 mm) in an axial image, in which standard deviation (SD) of grayscale values was measured and contrast-to-noise ratio (CNR) was calculated. Analysis of variance was used to compare the data.
Results:
Overall, in cases where the artifact was pronounced, MAR was efficient in reducing SD values. MAR also improved the CNR of ProMax images, but did not affect the Picasso images. Additionally, the higher was the kVp, the lower was the SD value and the higher was the CNR in both machines.
Conclusion:
In both machines, increasing kVp and MAR are effective in decreasing the CBCT artifacts in all their magnitude when they are pronounced. Therefore, the professionals should choose one of those options or even both considering the purpose of the CBCT imaging and radiation dose for the patient.
More Related Videos
05:37Cone Beam Intraoperative Computed Tomography-based Image Guidance for Minimally Invasive Transforaminal Interbody Fusion
Published on: August 6, 2019
10:23Author Spotlight: Three-Dimensional Cephalometric Landmark Annotation Demonstration on Human Cone Beam Computed Tomography Scans
Published on: September 8, 2023
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Orders of Magnitude
The order of magnitude is simply a way of rounding numbers consistently to the nearest power of 10. This makes doing rough mental math...
Factors Influencing Drug Absorption: Pharmaceutical Parameters
Imaging Studies III: Computed Tomography
Factors Influencing Drug Absorption: Anatomical Parameters
Factors Influencing Drug Absorption: Physicochemical Parameters
Enhanced drug absorption can be achieved by reducing particle sizes and increasing surface areas, thereby facilitating...