Hemodynamic performance within crossed stent grafts: computational and experimental study on the effect of cross
Ming Liu1, Anqiang Sun1,2, Xiaoyan Deng3,4
1Key Laboratory for Biomechanics and Mechanobiology of Ministry of Education, School of Biological Science & Medical Engineering, Beihang University, Beijing, 100083, China.
Background And Aims:
The crossed limbs stent graft technique is regularly employed to treat abdominal aortic aneurysm patients with unfavorable aneurysm necks or widely splayed common iliac arteries. This article numerically evaluates the hemodynamic performance of the crossed limbs strategy by analyzing numerical simulations and conducting experiments using two series of idealized bifurcated stent grafts with different cross angles and cross positions.
Results:
Results demonstrated that the absolute helicity at outlets decreased with increased cross angles and increased with decreased cross positions. The time-averaged wall shear stress remained approximately unchanged, whereas the oscillating shear index and relative resident time decreased slightly when the cross angle increased and cross position decreased in iliac grafts. Additionally, both numerical and in vitro experimental results indicate the displacement force acting on the stent graft gradually increased as cross angles increased and cross positions decreased. Results further indicated that strip areas with a high oscillating shear index and high relative resident time, which may be vulnerable to thrombosis formation, exist along the outer surface of the iliac artery grafts.
Conclusion:
Given this information, the optimal crossed limbs configuration may contain a small cross angle and low cross position; however, low cross positions may increase the risk of migration.
More Related Videos
06:55Fabrication of Small Caliber Stent-grafts Using Electrospinning and Balloon Expandable Bare Metal Stents
Published on: October 26, 2016
11:00Experimental Investigation of Secondary Flow Structures Downstream of a Model Type IV Stent Failure in a 180° Curved Artery Test Section
Published on: July 19, 2016
Related Concept Videos
Crossing Over
The homologous pairs of sister chromosomes—one from the maternal and one from the paternal genome—then begin to align alongside each other lengthwise, matching corresponding DNA positions in a process...
Crossing Over
Cross-Sectional Research
Monohybrid Crosses
Dihybrid Crosses
Test Cross