Development of a Benchtop Model for Evaluating the Compatibility of Wound Dressing Materials with Negative Pressure Wound Therapy Systems
Published on: May 2, 2025
MRI compatibility of silver based wound dressings
J Kevin Bailey1, Steffen Sammet2, Jason Overocker3
1Department of Surgery, Critical Care, Trauma and Burns, The Ohio State University, Columbus, OH, United States; Research Department, Shriners Hospitals for Children, Cincinnati, OH, United States.
Abstract:
As silver dressings gain more widespread use, it is more likely that patients with silver-based dressings will also undergo magnetic resonance imaging (MRI). In current practice, these dressings are removed prior to imaging due to concerns over heating and image distortion. As dressing changes can be painful, the need to remove dressings simply for MRI may increase pain and contribute to opioid dependency. To examine the need for dressing removal, American Society for Testing and Materials International standards for assessing device deflection and torque were performed on 5 silver containing and 3 non-silver control dressings. Magnetically induced heating and image distortion were examined in a porcine hind limb wound dressed with control and test dressings. The limb was scanned in a clinical high field 3T MRI scanner using a series of standard MRI sequences (Survey, T1-weighted SE, T1-weighted IR TSE, T2-weighted TSE, DUAL TSE, and FLAIR). Deflection and torsion were not detected in control or silver-based dressings. For all combinations of dressings and MRI scans, average heating was between 0-0.2°C. Additionally, dressings, in dry and hydrated forms, caused no image distortion in any MRI scan performed. Evaluation of MRI safety and compatibility revealed no concerns for safety or image distortion in any of the silver-containing wound dressings tested thus it would be acceptable to leave these dressings intact during MRI. The ability to leave dressings in place during imaging will provide a significant benefit to patient care by reducing pain associated with dressing removal.
Insights
Silver dressings are safe for magnetic resonance imaging (MRI) scans. Studies show these wound dressings do not cause heating or image distortion, reducing patient pain and opioid dependency.
Area of Science:
- Biomaterials Science
- Medical Imaging
- Wound Care
Background:
- Silver-based wound dressings are increasingly common.
- Current practice mandates their removal before magnetic resonance imaging (MRI) due to safety concerns.
- Dressing removal can cause pain and increase opioid use.
Purpose of the Study:
- To evaluate the MRI safety and compatibility of silver-containing wound dressings.
- To determine if silver dressings cause heating or image distortion during MRI.
- To assess the need for removing silver dressings before MRI scans.
Main Methods:
- Silver and control dressings were tested using American Society for Testing and Materials International standards for deflection and torque.
- Heating and image distortion were assessed in a porcine model using a 3T MRI scanner and various MRI sequences.
- Dry and hydrated dressings were evaluated.
Main Results:
- No deflection or torsion was detected in silver or control dressings.
- Magnetically induced heating was minimal (0-0.2°C) across all tested dressings and MRI sequences.
- No image distortion was observed with any silver-containing dressings, in either dry or hydrated states.
Conclusions:
- Silver-containing wound dressings are safe for use during MRI.
- These dressings do not pose a risk of heating or image artifact.
- Leaving silver dressings intact during MRI can reduce patient pain and potentially decrease opioid dependency.
More Related Videos
07:34Functional MRI in Conjunction with a Novel MRI-compatible Hand-induced Robotic Device to Evaluate Rehabilitation of Individuals Recovering from Hand Grip Deficits
Published on: November 23, 2019
07:22A Standardized Procedure of Dressing Management for Toxic Epidermal Necrolysis
Published on: March 14, 2025
Related Concept Videos
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Phases of Wound Repair
Formation of Blood Clot
In case of deep injuries, trauma to blood vessels results in blood loss. In the meantime, phospholipids released from the ruptured endothelial cellular membrane are converted into arachidonic...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Lewis Acids and Bases
A coordinate covalent bond (or dative bond) occurs when one of the atoms in the bond provides both bonding electrons. For example, a coordinate covalent bond occurs when a water molecule combines with a hydrogen ion to form a hydronium ion. A coordinate covalent bond also results when...
Base Excision Repair
The first step of...
Weak Base Solutions