Safety of orogastric tubes in foregut and bariatric surgery
Related Experiment Video
Updated: Feb 8, 2026

Author Spotlight: Recent Advancements in Reoperative Foregut Surgery
Published on: September 22, 2023
1Department of Surgery, Thomas Jefferson University Hospital, 1015 Walnut Street, Curtis Building, Suite 620, Philadelphia, PA, 19107, USA. Kulvir.Nandra@jefferson.edu.
Background:
Orogastric tubes have traditionally aided foregut procedures with sizing and organ protection. The rise of bariatric surgery has led to the creation of novel medical devices aimed at facilitating the laparoscopic sleeve gastrectomy. While approved by the FDA, the long-term safety profile of these devices in the general population is often unknown. This review looks at complications associated with novel Orogastric Tubes compared to the traditional bougie.
Methods:
We performed a review of the Food and Drug Administration's (FDA) Manufacturer and User Facility Device Experience (MAUDE) database for complications associated with the traditional bougie, Boehringer Labs ViSiGi 3D® and the Medtronic GastriSail™ since 2011. In addition, we looked for reported cases in the literature of complications with these devices.
Results:
Overall complication rates reported in the MAUDE database varied in number and severity. The bougie had seven reported complications, one of which was an organ perforation. The ViSiGi 3D® had zero reported complications. The GastriSail™ had 36 total reported complications with 17 perforations. A literature review shows that rates of bougie complications are extremely rare with no case reports or reviews of complications from the novel orogastric tubes.
Conclusions:
The complication rates between the traditional bougie and novel devices vary in number and severity, with the GastriSail™ having the highest reported complication rate. Despite rigorous testing for FDA approval, ongoing research into performance of new medical devices in the general population remains important.
More Related Videos
06:12Inducing Acute Liver Injury in Rats via Carbon Tetrachloride CCl4 Exposure Through an Orogastric Tube
Published on: April 28, 2020
11:46Investigation of the Electrophysiological and Thermographic Safety Parameters of Surgical Energy Devices During Thyroid and Parathyroid Surgery in a Porcine Model
Published on: October 13, 2022
Related Concept Videos
Survey Safety
Enteral Nutrition I: Orogastric and Nasogastric Feeding
Orogastric (OG) and nasogastric (NG) feeding are two standard methods used for enteral nutrition. Enteral nutrition is often preferred over...
Uterine Tubes
Household Wiring And Electrical Safety
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals....
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...