Periradicular infiltration of the lumbar spine: is iterative reconstruction software necessary to establish
Fabian Henry Jürgen Elsholtz1, Lars-Arne Schaafs2, Christoph Erxleben2
1Klinik und Hochschulambulanz für Radiologie, Charité- Universitätsmedizin Berlin, Campus Benjamin Franklin, Hindenburgdamm 30, 12203, Berlin, Germany. fabian.elsholtz@charite.de.
Purpose:
Computed tomography (CT)-guided periradicular infiltration therapy has emerged as an effective treatment option for patients with low back pain. Concern about radiation exposure requires approaches allowing significant dose reduction. The purpose of this study is to evaluate the need for iterative reconstruction software in CT-guided periradicular infiltration therapy with an ultra-low-dose protocol.
Materials And Methods:
One hundred patients underwent CT-guided periradicular infiltration therapy of the lumbar spine using an ultra-low-dose protocol with adaptive iterative dose reduction 3D (AIDR 3D) for image reconstruction. In addition, images were reconstructed with filtered back-projection (FBP). Four experienced raters evaluated both reconstruction types for conspicuity of anatomical and instrumental features important for ensuring safe patient treatment. Image noise was measured as a quantitative marker of image quality.
Results:
Interrater agreement was good for both AIDR 3D (Kendall's W = 0.83) and FBP (0.78) reconstructions. Readers assigned the same scores for all features and both reconstruction algorithms in 81.3% of cases. Image noise was significantly lower (average SD of 60.07 vs. 99.54, p < 0.05) for AIDR 3D-reconstructed images.
Conclusion:
Although it significantly lowers image noise, iterative reconstruction software is not mandatory to achieve adequate image quality with an ultra-low-dose CT protocol for guiding periradicular infiltration therapy of the lumbar spine.
Related Concept Videos
Qualitative Analysis
For instance, group IV...
Qualitative Analysis
There are two main approaches to qualitative analysis:...
Dose Size and Dosing Frequency: Determination Methods
Determination of Multiple Dosing Parameters: Loading and Maintenance Doses
Rational Dosage Regimen: Maintenance Dose and Loading Dose
In most cases, drugs are administered repetitively or infused continuously to maintain a steady-state concentration in the body. At a steady...
Reconstruction of Signal using Interpolation