Value of high-resolution mapping in optimizing cryoballoon ablation of atrial fibrillation
Giulio Conte1, Kyoko Soejima2, Carlo de Asmundis3
1Division of Cardiology, Cardiocentro Ticino, Lugano, Switzerland.
Background:
Unrecognized incomplete pulmonary vein isolation (PVI), as opposed to post-PVI pulmonary vein reconnection, may be responsible for clinical recurrences of atrial fibrillation (AF). To date, no data are available on the use of high-resolution mapping (HRM) during cryoballoon (CB) ablation for AF as the index procedure. The aims of this study were: - to assess the value of using a HRM system during CB ablation procedures in terms of ability in acutely detecting incomplete CB lesions; - to compare the 8-pole circular mapping catheter (CMC, Achieve) and the 64-pole mini-basket catheter (Orion) with respect to pulmonary vein (PV) signals detection at baseline and after CB ablation; - to characterize the extension of the lesion produced by CB ablation by means of high-density voltage mapping.
Methods:
Consecutive patients with drug-resistant paroxysmal or early-persistent AF undergoing CB ablation as the index procedure, assisted by a HRM system, were retrospectively included in this study.
Results:
A total of 33 patients (25 males; mean age: 59 ± 18 years, 28 paroxysmal AF) were included. At baseline, CMC catheter revealed PV activity in 102 PVs (77%), while the Orion documented PV signals in all veins (100%). Failure of complete CB-PVI was more frequently revealed by atrial re-mapping with the Orion as compared to the Achieve catheter (24% vs 0%, p < 0.05). A repeat ablation was performed in 8 patients (24%). In 9% of cases, the Orion catheter detected far-field signals originating from the right atrium. Quantitative assessment of the created lesion revealed a significant reduction of the left atrial area having voltage >0.5 mV. A total of 29 patients (88%) remained free of symptomatic AF during a mean follow-up of 13.2 ± 3.7 months.
Conclusion:
Atrial re-mapping after CB ablation by means of a HRM system improves the detection of areas of incomplete ablation, characterizes the extension of the cryo-ablated tissue and can identify abolishment of potential non-PVI related sources of AF.
More Related Videos
08:00Reduction of Iatrogenic Atrial Septal Defects with an Anterior and Inferior Transseptal Puncture Site when Operating the Cryoballoon Ablation Catheter
Published on: June 15, 2015
28:13Catheter Ablation in Combination With Left Atrial Appendage Closure for Atrial Fibrillation
Published on: February 26, 2013
Related Concept Videos
Amyloid Fibrils
Amyloid deposits were observed as early as 1639 in the liver and the spleen. In 1854, Rudolph Virchow performed iodine staining,...
Amyloid Fibrils
Fibril-associated Collagen
For example, the type II collagen fibrils in cartilage have covalently bound type IX fibril-associated collagens at regular intervals. Other types of fibril-associated collagens are...
Optimal Foraging
Optimization Problems
Microtubule Associated Proteins (MAPs)