Related Experiment Video
Updated: Feb 8, 2026

Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
Relative Biological Effectiveness Uncertainties and Implications for Beam Arrangements and Dose Constraints in Proton
Harald Paganetti1, Drosoula Giantsoudi1
1Department of Radiation Oncology, Massachusetts General Hospital and Harvard Medical School, Boston, MA.
Abstract:
Current clinical implementation of proton radiation therapy assumes a constant relative biological effectiveness (RBE) value of 1.1 throughout the treatment field, for both the target and organs at risks. Although few in vivo clinical data suggest that this approximation is clinically significant, in vitro studies demonstrate the dependency of RBE on dose, fractionation, proton energy, and linear energy transfer, as well as patient radiosensitivity and definition of endpoint. This article provides a brief review on the principles and individual factors contributing to RBE uncertainties, with emphasis on clinical practice. Clinical considerations regarding the effect of RBE uncertainties and implications for beam arrangements in proton therapy treatment planning are discussed through clinical examples for treatments of prostate cancer and posterior fossa tumors as well as craniospinal irradiation for medulloblastoma. Approaches on biological optimization in proton therapy are presented, including a discussion on linear energy transfer-based optimization as an alternative method for biological optimization and its implementation both in multicriteria optimization and inverse optimization modules.
More Related Videos
08:16Method for the Assessment of Effects of a Range of Wavelengths and Intensities of Red/near-infrared Light Therapy on Oxidative Stress In Vitro
Published on: March 21, 2015
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
The Uncertainty Principle
Biological Effects of Radiation
Insulin: Dosing Regimen and Adverse Effects
The basal dose constitutes about 40%-50% of the total daily dose, with the rest as premeal insulin. The mealtime insulin dose should mirror...
Uncertainty in Measurement: Reading Instruments
Uncertainty: Overview
Constraints and Statical Determinacy