Brandon C Perry1, Christopher R Ingraham2, Brent K Stewart1
1Department of Radiology, University of Washington, 1959 NE Pacific Street, Box 357987, Seattle, WA 98195.
Radiation dose monitoring software effectively tracks high-dose procedures in interventional radiology (IR) and neuro-interventional radiology (NIR). Patient-reported radiation-induced tissue reactions were uncommon, with temporary skin/hair injuries in 10.2% of high-dose NIR cases.
Area of Science:
- Medical Physics
- Radiology
- Interventional Radiology
Background:
- Fluoroscopically guided interventional radiology (IR) and neuro-interventional radiology (NIR) procedures can involve significant radiation doses.
- Effective monitoring and clinical follow-up are crucial for managing potential radiation-induced tissue reactions.
Purpose of the Study:
- To assess the implementation of radiation dose monitoring software in IR and NIR procedures.
- To establish a process for clinical follow-up and documentation of high-dose cases.
- To quantify patient-reported radiation-induced tissue reactions.
Main Methods:
- Web-based radiation dose monitoring software was implemented to flag procedures with reference point air kerma (Ka,r) > 5000 mGy.
- A retrospective chart review was conducted for high-dose procedures over a 1-year period.
- Patient-reported tissue reactions were evaluated for all flagged cases.
Main Results:
- The software successfully transferred dose data for 93.9% of procedures.
- Ten of 368 IR procedures (2.7%) and 52 of 1025 NIR procedures (5.1%) exceeded 5000 mGy.
- Five of 49 NIR patients (10.2%) reported temporary skin/hair injuries; no reactions were reported in IR cases.
Conclusions:
- Radiation dose monitoring software can be successfully implemented in interventional radiology settings.
- Radiation-induced tissue reactions are relatively uncommon following fluoroscopically guided procedures.
- A systematic approach to monitoring and follow-up is effective in managing high-dose cases.
Related Concept Videos
Biological Effects of Radiation
Nursing Interventions I: Taxonomy of Nursing Interventions
A nursing intervention is a treatment or action based on scientific concepts and knowledge from the nursing, behavioral, and physical sciences. Identifying and prioritizing nursing interventions based on the desired outcome...
Nursing Interventions II: Selecting and Classifying the Nursing Interventions
Radiological Investigation I: X-ray and CT
Radiation: Applications
The average...
Absorption of Radiation