Mechanomyography for Intraoperative Assessment of Cortical Breach During Instrumented Spine Surgery
Hesham Mostafa Zakaria1, Kelly M Tundo1, Carla Sandles1
1Department of Neurosurgery, Henry Ford Hospital, Detroit, Michigan, USA.
Objective:
We sought to determine the utility of mechanomyography (MMG) in detecting and preventing pedicle breach in instrumented lumbar spine surgery.
Methods:
In a prospective nonrandomized trial without controls, we selected consecutive patients to undergo intraoperative MMG during instrumented lumbar spine surgery. MMG testing was performed at the original pilot hole, after tapping, and after screw placement, with the minimum current to elicit a recorded MMG response. All patients underwent a postoperative computed tomography scan, and a single radiologist interpreted each pedicle to identify breach. Chi-square test was used to compare patients with and without breaches. Two sample Student's t-tests were used to compare changes in functional outcomes. Sensitivity and specificity of MMG were computed using receiver operating characteristic curve analysis.
Results:
There were 122 consecutive instrumented lumbar surgery patients enrolled, with a total of 890 lumbar pedicle screws tested with MMG. The medial or inferior breach rate was 2.25%, with no statistically significant difference in Oswestry Disability Index or visual analog scale between patients who breached and who did not. For the MMG measurement from the original pilot hole, the area under the receiver operating characteristic was 0.835; the maximum combination of sensitivity (80.42%) and specificity (80.6%) was found using MMG current ≤12 mA. We found that an MMG cutoff of >12 mA resulted in a 99.5% likelihood of no medial or inferior breach.
Conclusions:
MMG can be safely used during instrumented lumbar spine surgery. A cutoff value of >12 mA for MMG can accurately predict and prevent medial and inferior pedicle screw breach.
Insights
Mechanomyography (MMG) can detect pedicle breaches during lumbar spine surgery. An MMG cutoff of >12 mA accurately predicts and prevents medial or inferior breaches, improving surgical safety.
Area of Science:
- Spine Surgery
- Orthopedics
- Neurosurgery
Background:
- Pedicle breach is a complication in instrumented lumbar spine surgery.
- Accurate screw placement is crucial for surgical success and patient outcomes.
Purpose of the Study:
- To evaluate the effectiveness of mechanomyography (MMG) in identifying and preventing pedicle breaches during lumbar spine surgery.
- To establish a reliable MMG threshold for detecting breaches.
Main Methods:
- Prospective study of 122 patients undergoing instrumented lumbar spine surgery.
- Intraoperative MMG used to measure minimum current for response at pilot hole, after tapping, and after screw placement.
- Postoperative CT scans analyzed for pedicle breaches; MMG sensitivity and specificity calculated.
Main Results:
- A total of 890 pedicle screws were tested; the medial or inferior breach rate was 2.25%.
- MMG demonstrated an area under the receiver operating characteristic curve of 0.835.
- An MMG cutoff of >12 mA predicted no medial or inferior breach with 99.5% likelihood, achieving 80.42% sensitivity and 80.6% specificity.
Conclusions:
- Mechanomyography is a safe and effective tool for intraoperative use in instrumented lumbar spine surgery.
- An MMG current cutoff value greater than 12 mA accurately predicts and helps prevent medial and inferior pedicle screw breaches.
Related Concept Videos
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Instrumentation Amplifier
To overcome this challenge, an ECG machine utilizes an instrumentation amplifier. This specialized amplifier is...
Instrument Transformers
Uncertainty in Measurement: Reading Instruments
Capillary Electrophoresis: Instrumentation
Fluorescence and Phosphorescence: Instrumentation