Automated 90Sr Separation and Preconcentration in a Lab-on-Valve System at Ppq Level
Published on: June 6, 2018
Automated 90Sr Separation and Preconcentration in a Lab-on-Valve System at Ppq Level
Victor Vicente Vilas1, Sylvain Millet1, Miguel Sandow1
1Directorate G - Nuclear Safety & Security, Unit G.III. 8 - Waste Management, European Commission, DG Joint Research Centre - JRC.
Abstract:
A quick, automated and portable system for the separation and determination of radiostrontium in aqueous samples, using Sr-resin and multi sequential flow injection analysis, has been developed. The concentrations of radioactive strontium were determined by flow scintillation counting, allowing for on-line and also on-site determination. The proposed system can determine radioactive strontium at industrial relevant levels without further modification using overall analysis time of less than 10 min per aqueous sample. The limit of the detection is 320 fg·g-1 (1.7 Bq/g).
Related Concept Videos
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...