Related Experiment Video
Updated: Feb 8, 2026

A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Subcutaneous implantable cardioverter-defibrillator: is it ready for use in children and young adults? A
Massimo Stefano Silvetti1, Vincenzo Pazzano1, Letizia Verticelli1
1Pediatric Cardiology and Cardiac Arrhythmias Unit, Department of Pediatric Cardiology and Cardiac Surgery, Bambino Gesù Children's Hospital and Research Institute, Palidoro, Via Torre di Palidoro 1, Palidoro-Fiumicino, Rome, Italy.
Aims:
Use of the subcutaneous implantable cardioverter-defibrillator (S-ICD) to prevent sudden cardiac death is increasing. Few data exist on S-ICD in young patients. We reviewed our single-centre experience in order to assess the effectiveness and safety of S-ICD implantation, and to identify potential factors associated with complications.
Methods And Results:
Observational, non-randomized, standard-of-care study on S-ICD implantation/follow-up in young patients with inherited arrhythmias (IA), cardiomyopathies, and congenital heart defects (CHD). Fifteen patients (6 CHD, 8 cardiomyopathies, and 1 IA), median age 15 years (25th-75th centiles, 14-28), 10 of them <18 years, with body mass index (BMI) 22.6 ± 3.4, underwent S-ICD implantation (primary prevention 93%). The first six patients underwent a standard implantation procedure (three surgical incisions), the following nine a two-incision procedure. No intraoperative complications occurred. Over 12 (7-24) months follow-up, two patients received appropriate shocks and one (7%) inappropriate shock. Four patients had device-related complications requiring surgical intervention: three skin erosions at the superior parasternal incision, one pocket infection. A higher risk of complications was seen in patients who underwent standard procedures [hazard ratio (HR) 4.98, 95% confidence interval (CI) 1.53-47.36; P = 0.0482] and those with BMI <20 (HR 9.44, 95% CI 0.95-93.23; P = 0.0169).
Conclusion:
S-ICD implantation was safe and effective in young patients with low rates of inappropriate shock. However, the risk of device-related complications during follow-up remains substantial in patients with low BMI. To prevent possible complications, a two-incision technique should be preferred.