Intracardiac Flow at 4D CT: Comparison with 4D Flow MRI
Jonas Lantz1, Vikas Gupta1, Lilian Henriksson1
1From the Division of Cardiovascular Medicine, Department of Medical and Health Sciences (J.L., V.G., T.E.), Center for Medical Image Science and Visualization (CMIV) (J.L., V.G., L.H., M.K., A.P., C.J.C., T.E.), Division of Radiology, Department of Medical and Health Sciences (L.H., A.P.), Division of Applied Thermodynamics and Fluid Mechanics, Department of Management and Engineering (M.K.), and Department of Clinical Physiology, Department of Medical and Health Sciences (C.J.C.), Linköping University, SE-58183 Linköping, Sweden. From the 2017 RSNA Annual Meeting.
Insights
Four-dimensional flow CT accurately assesses intracardiac blood flow patterns, showing results comparable to four-dimensional flow MRI. This CT technique offers a reliable alternative for evaluating complex blood flow dynamics within the heart.
Area of Science:
- Cardiovascular Imaging
- Medical Physics
- Biomedical Engineering
Background:
- Four-dimensional (4D) flow MRI is the reference standard for assessing intracardiac blood flow.
- Developing advanced CT techniques for similar assessments is crucial for broader clinical application.
Purpose of the Study:
- To compare the accuracy of four-dimensional (4D) flow CT with 4D flow MRI in evaluating intracardiac blood flow patterns.
- To determine the quantitative and qualitative similarities between the two imaging modalities.
Main Methods:
- Prospective study of 12 participants comparing 4D flow CT and 4D flow MRI data.
- Quantification of flow patterns, rates, stroke volume, kinetic energy, and flow components for both techniques.
- Comparison using linear regression and statistical analysis for correlation and variability.
Main Results:
- 4D flow CT demonstrated qualitatively similar cardiac flow patterns to 4D flow MRI, confirmed by observer grading.
- High correlations were found for peak flow rate (r=0.98), stroke volume (r=0.81), and kinetic energy at peak systole (r=0.95).
- Significant correlations were also observed for direct (r=0.93) and residual (r=0.87) flow components.
Conclusions:
- Four-dimensional flow CT provides a reliable assessment of intracardiac blood flow patterns.
- The quantitative and qualitative results from 4D flow CT are comparable to the established 4D flow MRI standard.
- 4D flow CT presents a promising alternative for detailed intracardiac hemodynamics evaluation.
Abstract:
Purpose To investigate four-dimensional (4D) flow CT for the assessment of intracardiac blood flow patterns as compared with 4D flow MRI. Materials and Methods This prospective study acquired coronary CT angiography and 4D flow MRI data between February and December 2016 in a cohort of 12 participants (age range, 36-74 years; mean age, 57 years; seven men [age range, 36-74 years; mean age, 57 years] and five women [age range, 52-73 years; mean age, 64 years]). Flow simulations based solely on CT-derived cardiac anatomy were assessed together with 4D flow MRI measurements. Flow patterns, flow rates, stroke volume, kinetic energy, and flow components were quantified for both techniques and were compared by using linear regression. Results Cardiac flow patterns obtained by using 4D flow CT were qualitatively similar to 4D flow MRI measurements, as graded by three independent observers. The Cohen κ score was used to assess intraobserver variability (0.83, 0.79, and 0.70) and a paired Wilcoxon rank-sum test showed no significant change (P > .05) between gradings. Peak flow rate and stroke volumes between 4D flow MRI measurements and 4D flow CT measurements had high correlation (r = 0.98 and r = 0.81, respectively; P < .05 for both). Integrated kinetic energy quantified at peak systole correlated well (r = 0.95, P < .05), while kinetic energy levels at early and late filling showed no correlation. Flow component analysis showed high correlation for the direct and residual components, respectively (r = 0.93, P < .05 and r = 0.87, P < .05), while the retained and delayed components showed no correlation. Conclusion Four-dimensional flow CT produced qualitatively and quantitatively similar intracardiac blood flow patterns compared with the current reference standard, four-dimensional flow MRI. © RSNA, 2018 Online supplemental material is available for this article.
More Related Videos
07:07Author Spotlight: High-Resolution 4D Light-Sheet Imaging and Virtual Reality in Zebrafish for Single-Cell Analysis of Heart Function
Published on: January 5, 2024
08:384D Microscopy: Unraveling Caenorhabditis elegans Embryonic Development Using Nomarski Microscopy
Published on: October 8, 2020
Related Concept Videos
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Gene Flow
Flow Cytometry
In...
Flow Sheet
Here's a closer look at the examples of flowsheets commonly used by nurses:
Graphic Sheet Documentation:
Flow Table Test
Concrete is placed within a truncated cone mold that is 8 inches high with an 8-inch base diameter and a 5-inch top diameter. The...
Irrotational Flow