Field evaluation of onsite near real-time monitors for surface contamination by 5-fluorouracil
Jerome P Smith1, Deborah Sammons1, Shirley Robertson1
1NIOSH, Cincinnati, OH, USA.
Objectives:
In order to produce near real-time onsite results to detect surface contamination by antineoplastic drugs, the National Institute for Occupational Safety and Health developed monitors for 5-fluorouracil, which use surface wiping and lateral flow immunoassay for measurement. The monitors were tested in the laboratory to assess the sensitivity of detection on laboratory-produced contaminated surfaces. A field evaluation to assess the capability of the monitors to make measurements in healthcare workplaces was carried out in collaboration with a medical device company and the results are presented in this report.
Methods:
The 5-fluorouracil monitor was evaluated in areas where oncology drugs were prepared and administered to patients at five different hospitals. The levels of contamination measured with the monitors were compared to levels measured with liquid chromatography-tandem mass spectrometry.
Results:
The 5-fluorouracil values measured with the liquid chromatography-tandem mass spectrometry ranged from 0 to over 200,000 ng/100 cm2. Measurements by the 5-fluorouracil monitors in the range 10-100 ng/100 cm2 correlated with the liquid chromatography-tandem mass spectrometry. Receiver operating characteristic curves developed for the data indicated that a positive limit of 22 ng/100 cm2 would give an acceptable level of false-positives while retaining most true-positive samples. If the liquid chromatography-tandem mass spectrometry measured greater than 100 ng/100 cm2, then the monitors also measured levels greater than 100 ng/100 cm2 for the majority of samples.
Conclusion:
The data indicate that there are many areas in hospitals that are contaminated with 5-fluorouracil and the monitors will be useful in identifying this contamination.
Related Concept Videos
Equipotential Surfaces and Field Lines
Electric Field at the Surface of a Conductor
In the 19th century, Michael Faraday conducted the famous ice pail experiment to prove that the charges always reside on the surface of a conductor. The experimental set-up consists of a conducting uncharged container mounted on an insulating stand. The outer surface of the container is...
Real Time RT-PCR
The real-time quantification of the number of amplified products is...
Contaminants and Errors
Another key consideration is determining the appropriate number of samples required to...
5-Number Summary
In a box plot, the minimum and maximum data values represent the lower and upper whiskers in the graph, and the median is designated as the center of the box in the chart. The first quartile and third...
Surface Tension and Surface Energy
Consider a beaker filled with liquid. The bulk molecules in the liquid experience equal attractive forces on all sides with the surrounding molecules. However, the surface molecules experience a net attractive force downward due to the bulk molecules. The surface of the liquid behaves like a stretched membrane,...