Related Experiment Video
Updated: Feb 8, 2026

Osmotic Pump-based Drug-delivery for In Vivo Remyelination Research on the Central Nervous System
Published on: December 17, 2021
Prevalence of wireless smart-pump drug library update delays
Poching DeLaurentis1, Kang-Yu Hsu2, Yuval Bitan3
1Regenstrief Center for Healthcare Engineering, Purdue University, West Lafayette, IN poching@purdue.edu.
Purpose:
Results of a study to estimate the prevalence and severity of delays in wireless updates of smart-pump drug libraries across a large group of U.S. hospitals are reported.
Methods:
A prolonged smart-pump drug library update may result in patient harm if a pump is programmed with an incorrect limit setting at the time of drug administration. A retrospective study was conducted using smart-pump alert data extracted from the Regenstrief National Center for Medical Device Informatics (REMEDI) database. The study sample consisted of 49 hospitals in 5 states across the Midwest and Kentucky operated by 12 health systems; all the facilities used a specific brand of smart pump (BD Alaris, Beckton, Dickinson and Company) capable of generating alert data and had consistently contributed alert data to the REMEDI database over a 2-year period. An update delay was defined as the interval from the time a drug library version was replaced to the time of the last infusion alert triggered by the previous version during the study period.
Results:
Of the 12 health systems, 11 were found to have had drug library update delays during the study period, with delay medians ranging from 22 to 192 days. The overall delay minimum and maximum durations were 0 and 661 days.
Conclusion:
Substantial delays in completion of wireless updates of smart-pump drug libraries were common across a group of hospitals of various sizes.
Related Concept Videos
Prevalence and Incidence
Prevalence indicates the proportion of individuals in a population who have a specific disease or health...
ATP Driven Pumps III: V-type Pumps
The peripheral or cytosolic V1 domain with eight subunits is involved in ATP hydrolysis. The integral or transmembrane V0 domain containing at least five subunits...
ATP Driven Pumps II: P-type Pumps
A typical P-type pump has three cytosolic domains: nucleotide-binding (N), phosphorylation (P), and activator (A) domains. These domains are connected to the membrane-spanning helices by short amino acid segments. ATP hydrolysis and covalent phosphoenzyme intermediate formation are crucial parts of the catalytic cycle. At the highly...
Acid Suppressive Drugs for Peptic Ulcer Disease: Proton Pump Inhibitors
Gastric acid, a potent cocktail of hydrogen and chloride ions, is produced in specialized parietal cells within the...
Pumped Concrete
For direct-acting pumps, the concrete enters the pump via the inlet valve under the action of gravity and suction created by the movement of the piston. This concrete is then forced into the pipeline and out through the outlet valve by the forward movement...
Refrigerators and Heat Pumps
A household refrigerator removes heat from...