Evaluating Targeting Accuracy in the Focal Plane for an Ultrasound-guided High-intensity Focused Ultrasound Phased-array System
Published on: March 6, 2019
Effects of breast structure on high-intensity focused ultrasound focal error
Kohei Okita1, Ryuta Narumi2, Takashi Azuma2
11Department of Mechanical Engineering, College of Industrial Technology, Nihon University, 1-2-1 Izumi-cho, Narashino, Chiba, 275-8575 Japan.
Background:
The development of imaging technologies and breast cancer screening allowed early detection of breast cancers. High-intensity focused ultrasound (HIFU) is a non-invasive cancer treatment, but the success of HIFU ablation was depending on the system type, imaging technique, ablation protocol, and patient selection. Therefore, we aimed to determine the relationship between breast tissue structure and focal error during breast cancer HIFU treatment.
Methods:
Numerical simulations of the breast cancer HIFU ablation were performed using digital breast phantoms constructed using the magnetic resonance imaging data obtained from 12 patients.
Results:
The focal shapes were distorted despite breast tissue representing soft tissue. Focal errors are caused by the complex distribution of fibroglandular tissue, and they depend on the target position and the arrangement of the transducer. We demonstrated that the focusing ratio increases with the decrease in the local acoustic inhomogeneity, implying that it may be used as an indicator to reduce the HIFU focal error depending on the breast structure.
Conclusions:
The obtained results demonstrated that the focal error observed during the breast cancer HIFU treatment is highly dependent on the structure of fibroglandular tissue. The optimal arrangement of the transducer to the target can be obtained by minimizing the local acoustic inhomogeneity before the breast cancer HIFU treatment.
More Related Videos
07:38Real-time Monitoring of High Intensity Focused Ultrasound HIFU Ablation of In Vitro Canine Livers Using Harmonic Motion Imaging for Focused Ultrasound HMIFU
Published on: November 3, 2015
08:19Controllable Nucleation of Cavitation from Plasmonic Gold Nanoparticles for Enhancing High Intensity Focused Ultrasound Applications
Published on: October 5, 2018
Related Concept Videos
Systematic Error: Methodological and Sampling Errors
Sampling errors originate from improper sampling methods or the wrong sample population. These errors can be minimized by refining the sampling strategy. Defective instruments or faulty calibrations are the sources of instrumental...
Fundamental Attribution Error
Random Error
Margin of Error
Sound Intensity
Standard Error of the Mean
The standard error of the mean is an example of a standard error. It is a unique standard deviation known as the standard deviation of the sampling distribution of the mean. The standard error of the mean is a statistic that calculates how correctly a sample distribution represents a...