Benchmarking of Siemens Linac in Electron Modes: 8-14 MeV Electron Beams
H Dowlatabadi1, A A Mowlavi2,3, M Ghorbani4
1Physics Department, School of Sciences, Payame Noor University of Mashhad, Mashhad, Iran.
Background:
Radiation therapy using electron beams is a promising method due to its physical dose distribution. Monte Carlo (MC) code is the best and most accurate technique for forespeaking the distribution of dose in radiation treatment of patients.
Material And Methods:
We report an MC simulation of a linac head and depth dose on central axis, along with profile calculations. The purpose of the present research is to carefully analyze the application of MC methods for the calculation of dosimetric parameters for electron beams with energies of 8-14 MeV at a Siemens Primus linac. The principal components of the linac head were simulated using MCNPX code for different applicators.
Results:
The consequences of measurements and simulations revealed a good agreement. Gamma index values were below 1 for most points, for all energy values and all applicators in percent depth dose and dose profile computations. A number of states exhibited rather large gamma indices; these points were located at the tail of the percent depth dose graph; these points were less used in in radiotherapy. In the dose profile graph, gamma indices of most parts were below 1. The discrepancies between the simulation results and measurements in terms of Zmax, R90, R80 and R50 were insignificant. The results of Monte Carlo simulations showed a good agreement with the measurements.
Conclusion:
The software can be used for simulating electron modes of a Siemens Primus linac when direct experimental measurements are not feasible.
More Related Videos
07:00Preparing Lamellae from Vitreous Biological Samples Using a Dual-Beam Scanning Electron Microscope for Cryo-Electron Tomography
Published on: August 5, 2021
10:25Single-Digit Nanometer Electron-Beam Lithography with an Aberration-Corrected Scanning Transmission Electron Microscope
Published on: September 14, 2018
Related Concept Videos
Electron Affinity
Electron Carriers
Over the many stages of cellular respiration, glucose breaks down into carbon dioxide and water. Electron carriers pick up electrons lost by glucose in these reactions, temporarily storing and releasing them into the electron...
Electron Behavior
Electrons are negatively charged subatomic particles that are attracted to an orbit around the positively-charged nucleus of an atom. They reside in locations that are associated with energy levels called shells and are further organized into sub-shells and orbitals within each shell.
Electrons Orbit the Nucleus
Electrons are found in specific locations outside of the nucleus. The shell in which an electron resides indicates the general energy level of the electron: those closer to the...
Electron Behavior
Electrons Orbit the Nucleus
Electrons are found in specific locations outside of the nucleus. The shell in which an electron resides indicates the general energy level of the electron: those closer to the nucleus have less energy,...
Electron Transport Chains
The ETC is comprised of...
Electron Orbital Model
The first shell is closest to the nucleus, and it has only one subshell with a single spherical orbital called the...