Comparative Evaluation of Foveal Avascular Zone on Two Optical Coherence Tomography Angiography Devices
Prachi A Dave1, Kunal K Dansingani2, Ayesha Jabeen3
1Dr. Shroff's Charity Eye Hospital, New Delhi, India.
Significance:
This study compares foveal avascular zone (FAZ) geometry in healthy eyes as imaged by two commercially available optical coherence tomography angiography (OCTA) devices. Foveal avascular zone measurements are repeatable and reproducible with each OCTA device, but interdevice agreement was poor. We provide conversion factors between devices.
Purpose:
The purpose of this study was to perform comparative evaluation of FAZ geometry in healthy eyes as imaged by two commercially available OCTA devices.
Methods:
Ninety-six eyes of 48 healthy subjects were imaged prospectively on each of two OCTA devices (DRI-OCT [Topcon Corporation, Tokyo, Japan]; Cirrus 5000 [Carl Zeiss Meditec Inc., Dublin, CA]). The FAZ was evaluated in the superficial capillary plexus layer of 6 × 6-mm foveal scans by two masked observers. Intraobserver and interobserver agreement was determined using intraclass correlation by using linear mixed models and Bland-Altman plots. K-means clustering was used to provide conversion values between two devices. Foveal avascular zone acircularity was calculated using scans from each device and compared.
Results:
The intraobserver repeatability for DRI-OCT was 0.95 (95% confidence interval [CI], 0.90 to 0.98) for observer A and 0.92 (95% CI, 0.83 to 0.96) for observer B. Intraobserver repeatability for Cirrus 5000 was 0.988 (95% CI, 0.972 to 0.995) for observer A and 0.993 (95% CI, 0.983 to 0.997) for observer B. The interobserver variability between observers A and B for DRI-OCT was 0.87 (0.73 to 0.94) and for Cirrus 5000 was 0.984 (95% CI, 0.964 to 0.993). Poor interdevice agreement (0.205 [95% CI, -0.202 to 0.554]) was noted, and conversion formulas were devised to convert FAZ area measurements from one device to another. No significant correlation was found when comparing FAZ acircularity indices between devices (P = .39).
Conclusions:
Repeatable and reproducible FAZ area measurements were obtained with each respective OCTA device, but interdevice agreement was poor, yet quantifiable and systematic with calculable conversion factors between devices.
Related Concept Videos
Hybrid Zones
Zones of Protection
Protective zones are defined by closed dashed lines, containing one or more components. A key characteristic of these zones is the strategic placement of...
Transition Zone
Self-Evaluation: Self-Enhancement and Self-Verification
Comparing Copy Number Variations and SNPs
Copy number variations or CNVs are the structural variations that cover more than 1kb of DNA sequence. The single nucleotide polymorphism (SNP), on the other hand, is a single nucleotide change or a point mutation that is found in more than 1%...
Comparative Excretory Systems