Performance characterization of an integrated cone-beam CT system for dedicated gamma radiosurgery
Arman Sarfehnia1,2, Mark Ruschin1,2, Brige Chugh1,2
1Department of Radiation Oncology, University of Toronto, 150 College St., Toronto, ON, M5S-3E2, Canada.
Purpose:
This work describes the performance characterization of a cone-beam CT-guided radiosurgery device, the Gamma Knife® Icon™.
Methods:
The performance tests have been categorized into: (a) image quality and mechanical integrity; (b) image coregistration fidelity; (c) adaptive treatment delivery quality; (d) high definition motion management performance characterization; (e) software communication performance testing of the integrated cone-beam CT (CBCT) system.
Results:
All image quality performance characterization satisfied or exceeded manufacturer specifications. The image quality and mechanical stability of the CBCT system over a 3-month period was within tolerance with negligible (<0.1°) detector tilt angle. The CBCT definition of the stereotactic space had a measured average discrepancy of 0.15-0.16 mm in x, y, and z directions. On average, the high definition motion management system performance was within 0.05 mm with a residual offset of 0.15 mm when large displacements in a given direction were taken. The adaptive treatment delivery component as measured with CBCT coregistration of daily setups against reference setup images was accurate to within 0.2°. Comprehensive end-to-end testing showed a total uncertainty of better than 0.2 mm in positioning and 0.4% in dosimetry for treatment of centrally located lesions.
Conclusions:
A set of system performance characterization tests spanning all aspects of the Gamma Knife Icon are presented. Overall, the system performance was in line with manufacturer specifications.
More Related Videos
05:49Author Spotlight: Advancing CBCT and Digital Dental Image Integration with AI-Assisted Digitization
Published on: February 23, 2024
09:49Dual-phase Cone-beam Computed Tomography to See, Reach, and Treat Hepatocellular Carcinoma during Drug-eluting Beads Transarterial Chemo-embolization
Published on: December 2, 2013
Related Concept Videos
Beams
Based on geometry, beams can be straight, tapered, or curved. Straight beams are the most common type and have a constant cross-section throughout their length. Tapered beams, on the other hand, have a varying cross-section along...
Deflection of a Beam
Singularity functions, described in an earlier lesson, are powerful mathematical tools that represent discontinuities within a function commonly encountered in structural loading...
Prismatic Beams: Problem Solving
The design begins with analyzing the beam as a free body to identify moments and force balances, thereby determining support reactions. Next, the...
Principal Stresses in a Beam
Analyzing principal stresses is crucial, especially in...
Beams with Symmetric Loadings
The M/EI...
Integration by Parts: Indefinite Integrals