An in vitro evaluation of turbulence after transcatheter aortic valve implantation
Hoda Hatoum1, Atieh Yousefi1, Scott Lilly2
1Department of Biomedical Engineering, The Ohio State University, Columbus, Ohio.
Background:
This study aimed at assessment of post-transcatheter aortic valve (TAV) replacement hemodynamics and turbulence when a same-size SAPIEN 3 (Edwards Lifesciences Corp, Irvine, Calif) and Medtronic Evolut (Minneapolis, Minn) were implanted in a rigid aortic root with physiological dimensions and in a representative root with calcific leaflets obtained from patient computed tomography scans.
Methods:
TAV hemodynamics were studied by placing a SAPIEN 3 26-mm and an Evolut 26-mm in rigid aortic roots and representative root with calcific leaflets under physiological conditions. Hemodynamics were assessed using high-fidelity particle image velocimetry and high-speed imaging. Transvalvular pressure gradients (PGs), pinwheeling indices, and Reynolds shear stress (RSS) were calculated.
Results:
(1) PGs obtained with the Evolut and the SAPIEN 3 were comparable among the different models (10.5 ± 0.15 mm Hg vs 7.76 ± 0.083 mm Hg in the rigid model along with 13.9 ± 0.19 mm Hg vs 5.0 ± 0.09 mm Hg in representative root with calcific leaflets obtained from patient computed tomography scans respectively); (2) more pinwheeling was found in the SAPIEN 3 than the Evolut (0.231 ± 0.057 vs 0.201 ± 0.05 in the representative root with calcific leaflets and 0.366 ± 0.067 vs 0.122 ± 0.045 in the rigid model); (3) higher rates of RSS were found in the Evolut (161.27 ± 3.45 vs 122.84 ± 1.76 Pa in representative root with calcific leaflets and 337.22 ± 7.05 vs 157.91 ± 1.80 Pa in rigid models). More lateral fluctuations were found in representative root with calcific leaflets.
Conclusions:
(1) Comparable PGs were found among the TAVs in different models; (2) pinwheeling indices were found to be different between both TAVs; (3) turbulence patterns among both TAVs translated according to RSS were different. Rigid aortic models yield more conservative estimates of turbulence; (4) both TAVs exhibit peak maximal RSS that exceeds platelet activation 100 Pa threshold limit.
More Related Videos
06:59Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
06:04Upper-extremity Approach for Secondary Access in Transfemoral Transcatheter Aortic Valve Implantation
Published on: August 8, 2025
Related Concept Videos
Laminar and Turbulent Flow
Turbulent Flow
Turbulent Flow: Problem Solving
Temperature is a key factor in CO2 solubility. In this case, the CO2 gas and the liquid are cooled to 20°C. Lower temperatures enhance...
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Mitral Valve Prolapse I: Introduction
Aortic Regurgitation I: Introduction