Radiation Exposure to the Personnel Performing Robotic Arm-Assisted Positron Emission Tomography/Computed
Tamanna Lakhanpal1, Bhagwant Rai Mittal1, Rajender Kumar1
1Department of Nuclear Medicine and Positron Emission Tomography/Computed Tomography, Postgraduate Institute of Medical Education and Research, Chandigarh, India.
Purpose:
The aim of the study was to estimate the radiation burden to personnel performing robotic arm-assisted positron emission tomography/computed tomography (PET/CT)-guided metabolic biopsies and to staff assisting the procedure in a PET/CT facility.
Materials And Methods:
We estimated the whole-body exposures to physicians and staff based on the dose rate measurement with an ionization chamber-based calibrated survey monitor and pocket dosimeters during the robotic arm-assisted 2-[F-18]-fluoro-2-deoxy-D-glucose PET/CT-guided biopsies from September 2016 to February 2017. In addition, we also noted the time to biopsy after injection and total biopsy time during which the staff was exposed to the radiation.
Results:
In this prospective study, PET/CT-guided biopsy was performed in 50 patients (males - 36, females - 14) with a mean age of 54 ± 15 years (range17-78 years). Of the 50 biopsy procedures, 18 were lung biopsies, 10 were bone biopsies, and 22 biopsies were from abdomen/pelvic lesions. The mean time taken for the procedure was 26 ± 11 min. The mean time elapsed between injection and procedure was 182 ± 85 min and mean injected activity of 138.38 ± 74 MBq. The mean whole-body exposure per procedure to an interventionist and an assistant was 1.88 ± 0.82 μSv and 1.04 ± 0.75 μSv, respectively. The mean exposure rates at 0-m and 1-m distance from the patient were 30.77 ± 14.61 μSv/h and 2.59 ± 1.49 μSv/h, respectively.
Conclusion:
We conclude that the interventionist received the highest mean effective dose as compared to the helping staffs. The occupational radiation exposure was found to be within the limits as specified by the regulatory authorities (International Commission on Radiological Protection-103), and PET/CT-guided biopsies were safe from radiation protection point of view.
More Related Videos
11:09High-Resolution Cardiac Positron Emission Tomography/Computed Tomography for Small Animals
Published on: December 16, 2022
05:32Multimodal Bioluminescent and Positronic-emission Tomography/Computational Tomography Imaging of Multiple Myeloma Bone Marrow Xenografts in NOG Mice
Published on: January 7, 2019
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Emission Spectra
Biological Effects of Radiation